FDA Adverse Event Malfunction Summary report: N

MEDLINE RENEWAL REPROCESSED ETHICON XCEL TROCAR

MDR report key: 6088136 · Received November 8, 2016

Report

Report Number
3032391-2016-00013
Event Type
Malfunction
Date Received
November 8, 2016
Date of Event
June 30, 2016
Report Date
November 8, 2016
Manufacturer
MEDLINE RENEWAL
Product Code
NLM
PMA / PMN Number
K012644
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IN QUESTION WAS RETURNED FOR EVALUATION. THE REPROCESSED ETHICON H12LP WAS REPROCESSED ONE TIME BY MEDLINE RENEWAL. UPON RECEIPT, WE CONFIRMED THAT THE LOCKING MECHANISM LOCATED ON THE ADJUSTABLE FIXATION PLUG WAS BROKEN. ALL OTHER COMPONENTS MET ESTABLISHED INSPECTION CRITERIA. IN ADDITION, WE PERFORMED A REVIEW OF THE DEVICE HISTORY RECORD FOR LOT NUMBER 302057. WE RECONFIRMED THAT ALL PROCESSES WERE CONDUCTED AS REQUIRED AND THAT THE DEVICE MET ALL THE INSPECTION REQUIREMENTS PRIOR TO PACKAGING AND RELEASE. UNFORTUNATELY, MEDLINE RENEWAL DOES NOT HAVE ENOUGH INFORMATION TO DETERMINE THE ROOT CAUSE OF THE FAILURE. EVEN THOUGH MEDLINE RENEWAL CONFIRMED THAT NO ADVERSE PATIENT CONSEQUENCE, PROCEDURAL DELAY, OR MEDICAL INTERVENTION WAS REQUIRED AS A RESULT OF THE DEVICE FAILURE, IN AN ABUNDANCE OF CAUTION MEDLINE RENEWAL IS FILING THIS MEDWATCH REPORT.

Description of Event or Problem · 1

MEDLINE RENEWAL RECEIVED A REPORT INDICATING THAT THE LOCK LOCATED ON THE ADJUSTABLE FIXATION PLUG OF A REPROCESSED ETHICON XCEL TROCAR, MODEL H12LP BROKE WHILE THE SCRUB TECH TRIED TO LOCK IT. ANOTHER DEVICE WAS READILY AVAILABLE TO COMPLETE THE PROCEDURE, THEREFORE THERE WAS NO ADVERSE PATIENT CONSEQUENCE, PROCEDURAL DELAY OR MEDICAL INTERVENTION REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
738244 MEDLINE RENEWAL REPROCESSED ETHICON XCEL TROCAR LAPAROSCOPIC/ENDOSCOPIC TROCAR AND CANNULA NLM MEDLINE RENEWAL H12LP 302057

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention