CORAIL2 ANTERIOR BROACH HANDLE
Report
- Report Number
- 1818910-2016-31125
- Event Type
- Malfunction
- Date Received
- November 8, 2016
- Date of Event
- November 3, 2016
- Report Date
- November 3, 2016
- Manufacturer
- DEPUY ORTHOPAEDICS, INC. 1818910
- Product Code
- HTQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
A FUNCTIONAL TEST WITH A MATING BROACH FOUND THE BROACH HANDLES LOCK ONTO THE BROACH; HOWEVER, THE HANDLES ARE LOOSE. THE ROOT CAUSE IS ATTRIBUTED TO MISUSE AND WEAR; DAMAGE TO THE CAM COMPONENTS. IT IS IMPORTANT TO ENSURE THAT THE ALIGNMENT FEATURES ON BOTH THE HANDLE AND BROACH ARE MATED CORRECTLY, AS PROPER LOCKING WILL EXTEND THE LIFE OF ALL BROACH HANDLES. IF THESE ORIENTATION FEATURES ARE NOT ALIGNED PROPERLY, THE CAM WILL LOCK INCORRECTLY ONTO THE BROACH. AFTER THIS TYPE OF DEFORMATION OCCURS, THE HANDLE WILL NO LONGER LOCK AS TIGHTLY ONTO THE BROACH AS WHEN FIRST DISTRIBUTED. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4). THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
THE BROACH HANDLES WERE FOUND BEFORE THE CASE IN THE SET AND NO LONGER WILL LOCK THE BROACH ON CORRECTLY. THE LOCKING MECHANISM DOES NOT WORK ANYMORE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 737269 | CORAIL2 ANTERIOR BROACH HANDLE | HIP INSTRUMENT/TRIAL | HTQ | DEPUY ORTHOPAEDICS, INC. 1818910 | NB47722 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |