FDA Adverse Event Malfunction Summary report: N

CORAIL2 ANTERIOR BROACH HANDLE

MDR report key: 6087973 · Received November 8, 2016

Report

Report Number
1818910-2016-31125
Event Type
Malfunction
Date Received
November 8, 2016
Date of Event
November 3, 2016
Report Date
November 3, 2016
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
HTQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A FUNCTIONAL TEST WITH A MATING BROACH FOUND THE BROACH HANDLES LOCK ONTO THE BROACH; HOWEVER, THE HANDLES ARE LOOSE. THE ROOT CAUSE IS ATTRIBUTED TO MISUSE AND WEAR; DAMAGE TO THE CAM COMPONENTS. IT IS IMPORTANT TO ENSURE THAT THE ALIGNMENT FEATURES ON BOTH THE HANDLE AND BROACH ARE MATED CORRECTLY, AS PROPER LOCKING WILL EXTEND THE LIFE OF ALL BROACH HANDLES. IF THESE ORIENTATION FEATURES ARE NOT ALIGNED PROPERLY, THE CAM WILL LOCK INCORRECTLY ONTO THE BROACH. AFTER THIS TYPE OF DEFORMATION OCCURS, THE HANDLE WILL NO LONGER LOCK AS TIGHTLY ONTO THE BROACH AS WHEN FIRST DISTRIBUTED. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4). THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE BROACH HANDLES WERE FOUND BEFORE THE CASE IN THE SET AND NO LONGER WILL LOCK THE BROACH ON CORRECTLY. THE LOCKING MECHANISM DOES NOT WORK ANYMORE.