FDA Adverse Event Malfunction Summary report: N

BOM 7MM EXTENDED LENGTH ENDOSCOPE

MDR report key: 6087563 · Received November 8, 2016

Report

Report Number
2242352-2016-01102
Event Type
Malfunction
Date Received
November 8, 2016
Report Date
October 11, 2016
Manufacturer
MAQUET CV
Product Code
GCJ
PMA / PMN Number
K014250
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS IS A REUSABLE OEM DEVICE; THEREFORE, A LOT HISTORY REVIEW WAS NOT APPLICABLE. THE CERTIFICATE OF CONFORMANCE (C OF C) WAS REVIEWED. THE VENDOR CERTIFIES THAT THIS DEVICE LOT CONFORMS TO ALL APPLICABLE PRODUCT SPECIFICATIONS. THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. SIGNS OF CLINICAL USE AND EVIDENCE OF BLOOD WERE OBSERVED. A VISUAL INSPECTION WAS CONDUCTED. NO VISUAL DEFECTS WERE OBSERVED. AN IMAGE QUALITY INSPECTION WAS PERFORMED WITH AN ENDOSCOPIC VIDEO IMAGING SYSTEM. A CLEAR IMAGE WAS UNABLE TO BE OBTAINED. CONDENSATION WAS OBSERVED ON THE INSIDE OF THE LENS WHEN THE DEVICE WAS OUT OF FOCUS. BASED ON THE RESULTS OF THE EVALUATION, THE REPORTED COMPLAINT WAS CONFIRMED FOR THE REPORTED FAILURE "POOR QUALITY IMAGE".

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR EVALUATION. WE ARE FOLLOWING UP WITH THE CUSTOMER FOR THE RETURN OF THE DEVICE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED.

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING PREPARATION FOR AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, BIOMED RECEIVED BOM 7MM EXTENDED LENGTH ENDOSCOPE S/N (B)(4) FOR REPAIR. IT WAS NOTICED THAT MOISTURE WAS IN THE LENS AND A BLURRY PICTURE ON MONITOR. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING PREPARATION FOR AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, BIOMED RECEIVED BOM 7MM EXTENDED LENGTH ENDOSCOPE S/N (B)(4) FOR REPAIR. IT WAS NOTICED THAT MOISTURE WAS IN THE LENS AND A BLURRY PICTURE ON MONITOR. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.