FDA Adverse Event Injury Summary report: N

PFCSIG CEM FEM STEM 5DG13X90MM

MDR report key: 6087318 · Received November 8, 2016

Report

Report Number
1818910-2016-31106
Event Type
Injury
Date Received
November 8, 2016
Date of Event
September 19, 2016
Report Date
November 2, 2016
Manufacturer
DEPUY-RAYNHAM 1219655
Product Code
NJL
PMA / PMN Number
P830055
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4). (THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE SUBMITTED DEVICE CONFIRMED FRACTURE DUE TO LOW-STRESS, HIGH-CYCLE FATIGUE. THERE WAS NO EVIDENCE OF INCLUSIONS OR OTHER MATERIAL DEFECTS THAT COULD HAVE CONTRIBUTED TO CRACK INITIATION OR PROPAGATION. REVIEW OF THE DEVICE HISTORY RECORDS DID NOT REVEAL ANY MANUFACTURING DEVIATIONS OR ANOMALIES. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. NO EVIDENCE WAS FOUND INDICATING PRODUCT ERROR WAS A CONTRIBUTING FACTOR TO THE REPORTED EVENT AND THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS PAIN, OSTEOLYSIS AND A BROKEN ADAPTOR BOLT AND LOOSE FEMORAL COMPONENT. LOOSENING OCCURRED AT THE BONE/CEMENT IMPLANT INTERFACE. CEMENT MANUFACTURED BY A COMPETITOR. UPDATE 11/2/2016: EXAMINATION OF THE SUBMITTED DEVICES FOUND THE STEM BOLT AND STEM FRACTURED. THE STEM HAS NOW BEEN ADDED AND REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
738114 PFCSIG CEM FEM STEM 5DG13X90MM KNEE OTHER IMPLANT NJL DEPUY-RAYNHAM 1219655 316349R

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention