PUREPOINT LASER
Report
- Report Number
- 2028159-2016-04992
- Event Type
- Malfunction
- Date Received
- November 7, 2016
- Report Date
- December 22, 2016
- Manufacturer
- ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
- Product Code
- HQF
- PMA / PMN Number
- K062624
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HU
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).
THE SYSTEM WAS EXAMINED AND THE REPORTED EVENT WAS REPLICATED. THE COMPANY REPRESENTATIVE RE-SEATED THE POWER CABLE AND RE-CALIBRATED THE SYSTEM AS A PREVENTIVE MEASURE. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. A REVIEW OF THE CUSTOMER¿S COMPLAINT HISTORY FOR THE LAST 24 MONTHS DID NOT SHOW ANY PREVIOUS COMPLAINTS OF THIS KIND AGAINST THE SYSTEM. THE SYSTEM WAS MANUFACTURED ON FEBRUARY 24, 2009. BASED ON QA ASSESSMENT, THE PRODUCT MET SPECIFICATIONS AT THE TIME OF RELEASE. THE SYSTEM WAS FOUND TO MEET SPECIFICATIONS. THEREFORE, THE ROOT CAUSE OF THE REPORTED EVENT CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).
A HEALTHCARE PROFESSIONAL REPORTED THAT THE LASER SHUT DOWN DURING CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 735578 | PUREPOINT LASER | LASER, OPHTHALMIC | HQF | ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |