FDA Adverse Event Malfunction Summary report: N

PUREPOINT LASER

MDR report key: 6081224 · Received November 7, 2016

Report

Report Number
2028159-2016-04998
Event Type
Malfunction
Date Received
November 7, 2016
Date of Event
October 11, 2016
Report Date
December 8, 2016
Manufacturer
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Product Code
HQF
PMA / PMN Number
K062624
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. (B)(4)

Additional Manufacturer Narrative · 1

THE SYSTEM WAS EXAMINED AND THE REPORTED EVENT WAS NOT REPLICATED. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. THE SYSTEM WAS MANUFACTURED ON AUGUST 12, 2009. BASED ON QA ASSESSMENT, THE PRODUCT MET SPECIFICATIONS AT THE TIME OF RELEASE. THE SYSTEM WAS FOUND TO MEET SPECIFICATIONS. THEREFORE, THE ROOT CAUSE OF THE REPORTED EVENT CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT DURING A PHOTOCOAGULATION PROCEDURE, THE LASER PRODUCED LOW ENERGY. THE CASE COULD NOT BE COMPLETED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
735516 PUREPOINT LASER LASER, OPHTHALMIC HQF ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER NA NA

Patients

Seq Age Sex Outcome Treatment
1 Other