PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2016-35465
- Event Type
- Malfunction
- Date Received
- November 4, 2016
- Date of Event
- November 16, 2013
- Report Date
- November 20, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- PATIENT
Narratives
THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, EXCESSIVE NO DELIVERY AND DISPLACEMENT TESTS. NO EXCESSIVE NO DELIVERY ALARMS NOTED. HOWEVER, AN UNEXPECTED MOTOR NOISE WAS NOTED DURING THE REWIND; PROBLEM ISOLATED TO MOTOR. IT ALSO HAD A CRACKED CASE AT LCD WINDOW CORNER, CRACKED RESERVOIR TUBE, CRACKED BATTERY TUBE THREADS, BROKEN RESERVOIR TUBE LIP AND SCRATCHED LCD WINDOW. NOTE: THIS IS A REMEDIATION MDR. MEDTRONIC DIABETES IMPLEMENTED REVISED MDR REPORTABILITY CRITERIA EFFECTIVE ON (B)(6) 2014. SUBSEQUENTLY, MEDTRONIC DIABETES CONDUCTED A ONE YEAR RETROSPECTIVE REVIEW OF COMPLAINTS. THIS EVENT WAS RETROSPECTIVELY IDENTIFIED TO BE REPORTABLE BASED ON THE REVISED MDR REPORTABILITY CRITERIA. (B)(4).
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP MAKES A GRINDING NOISE DURING REWIND AND ALARMED NO DELIVERY DURING PRIME. THE BLOOD GLUCOSE AT THE TIME OF THE INCIDENT WAS 300 MG/DL; TREATED WITH MANUAL INJECTIONS. DURING THE CALL, THE CUSTOMER CHANGED THE INFUSION SET AND RESERVOIR AND WAS ABLE TO PRIME. THE CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP HAS A CRACKED RESERVOIR COMPARTMENT AND SCREEN. THE DEVICE WAS RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 731159 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | UNOMEDICAL INSULIN INFUSION SET |