FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 6079565 · Received November 4, 2016

Report

Report Number
2032227-2016-35465
Event Type
Malfunction
Date Received
November 4, 2016
Date of Event
November 16, 2013
Report Date
November 20, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, EXCESSIVE NO DELIVERY AND DISPLACEMENT TESTS. NO EXCESSIVE NO DELIVERY ALARMS NOTED. HOWEVER, AN UNEXPECTED MOTOR NOISE WAS NOTED DURING THE REWIND; PROBLEM ISOLATED TO MOTOR. IT ALSO HAD A CRACKED CASE AT LCD WINDOW CORNER, CRACKED RESERVOIR TUBE, CRACKED BATTERY TUBE THREADS, BROKEN RESERVOIR TUBE LIP AND SCRATCHED LCD WINDOW. NOTE: THIS IS A REMEDIATION MDR. MEDTRONIC DIABETES IMPLEMENTED REVISED MDR REPORTABILITY CRITERIA EFFECTIVE ON (B)(6) 2014. SUBSEQUENTLY, MEDTRONIC DIABETES CONDUCTED A ONE YEAR RETROSPECTIVE REVIEW OF COMPLAINTS. THIS EVENT WAS RETROSPECTIVELY IDENTIFIED TO BE REPORTABLE BASED ON THE REVISED MDR REPORTABILITY CRITERIA. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP MAKES A GRINDING NOISE DURING REWIND AND ALARMED NO DELIVERY DURING PRIME. THE BLOOD GLUCOSE AT THE TIME OF THE INCIDENT WAS 300 MG/DL; TREATED WITH MANUAL INJECTIONS. DURING THE CALL, THE CUSTOMER CHANGED THE INFUSION SET AND RESERVOIR AND WAS ABLE TO PRIME. THE CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP HAS A CRACKED RESERVOIR COMPARTMENT AND SCREEN. THE DEVICE WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
731159 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAB

Patients

Seq Age Sex Outcome Treatment
1 48 YR UNOMEDICAL INSULIN INFUSION SET