FDA Adverse Event Death Summary report: N

ROTABLATOR ROTATIONAL ATHERECTOMY SYSTEM

MDR report key: 6073621 · Received November 2, 2016

Report

Report Number
2134265-2016-09973
Event Type
Death
Date Received
November 2, 2016
Date of Event
October 4, 2016
Report Date
October 5, 2016
Manufacturer
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
Product Code
MCX
PMA / PMN Number
P900056
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE BATCH NUMBER IS UNKNOWN AND THE MANUFACTURING RECORDS FOR THE COMPLAINT DEVICE COULD NOT BE REVIEWED. THE ROOT CAUSE IS ANTICIPATED PROCEDURAL COMPLICATIONS AS THIS EVENT IS A KNOWN PHYSIOLOGICAL EFFECT OF THE PROCEDURE AND IS NOTED WITHIN THE DFU. (B)(4).

Description of Event or Problem · 1

SAME CASE AS MDR ID: 2134265-2016-09559. IT WAS REPORTED THAT VESSEL PERFORATION AND PATIENT DEATH OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED MID LEFT ANTERIOR DESCENDING ARTERY (LAD). A 1.50MM ROTALINK PLUS AND A ROTAWIRE WERE USED TO ADVANCE AND TREAT THE TARGET LESION. A 330CM ROTAWIRE WAS USED TO ADVANCE TO THE TARGET LESION. DURING THE PROCEDURE, DIFFICULTY IN CROSSING THE LESION WAS ENCOUNTERED AND IT WAS NOTED THAT THE COIL PART OF THE WIRE WAS STRETCHED. THE WIRE WAS EXCHANGED WITH ANOTHER ROTAWIRE AND PERFORMED ABLATION FOR THREE TIMES AT 180,000RPM FOR ABOUT 60 SECONDS EACH. DURING THE THIRD ABLATION, IT WAS NOTED THAT THE CORONARY ARTERY WAS PERFORATED. AFTER THE PERFORATION, A NON BSC STENT WAS PLACED. HOWEVER, PERFORATION OCCURRED IN THE DISTAL OF THE STENT. THEN, A NON BSC BALLOON WAS USED IN ATTEMPT TO STOP THE BLEEDING. THEN, A SECOND STENT WAS PLACED IN FRONT OF THE FIRST STENT, BUT PERFORATION ALSO OCCURRED IN THE PROXIMAL OF THE SECOND STENT. THE PHYSICIAN ATTEMPTED TO STOP THE BLEEDING IN LAD AND COMPLETED THE PROCEDURE INSUFFICIENT. THE PROCEDURE TIME WAS 8 HOURS. AN INTRA-AORTIC BALLOON PUMP (IABP) AND PERCUTANEOUS CARDIOPULMONARY SYSTEMS (PCPS) WERE INSERTED TO THE PATIENT AND WAS MONITORED. THEN, THE PATIENT WAS PREPARED FOR TRANSFER TO ANOTHER FACILITY OF CARDIAC SURGERY. THE PATIENT DIED BEFORE THE TRANSFER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
724459 ROTABLATOR ROTATIONAL ATHERECTOMY SYSTEM CATHETER, CORONARY, ATHERECTOMY MCX BOSTON SCIENTIFIC - COSTA RICA (COYOL) UNK509

Patients

Seq Age Sex Outcome Treatment
1 67 YR Death