PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2016-34980
- Event Type
- Death
- Date Received
- November 2, 2016
- Date of Event
- September 29, 2016
- Report Date
- August 4, 2017
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSULIN PUMP PASSED ALL FUNCTIONAL TESTING INCLUDING THE IDLE CURRENT MEASUREMENT TEST, RUN CURRENT MEASUREMENT TEST, SELF -TEST, OFF NO POWER TEST, UNEXPECTED RESTART ERROR TEST, DISPLACEMENT TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, REWIND TEST, EXCESSIVE NO DELIVERY TEST AND DELIVERY ACCURACY TEST. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHED LCD WINDOW. DATA ANALYSIS THERE IS NO DATA LISTED FOR THE DATED OF (B)(6) 2016 DUE TO INSULIN PUMP RECEIVED WITHOUT BATTERY INSTALLED.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(4). PLEASE SEE MEDWATCH REPORTS 3004209178-2016-79888, 2032227-2016-23510, 2032227-2016-33035.
IT WAS REPORTED THAT THE CUSTOMER WAS IN A COMA IN THE HOSPITAL WITH BLOOD GLUCOSE OVER 1000 MG/DL. THE CALLER ASKED THE TERRITORY MANAGER WHEN THEY HAD LAST CONTACTED THE CUSTOMER SO THAT THEY COULD PUT TOGETHER A TIMELINE. THE TERRITORY MANAGER ADVISED THAT THE LAST TIME THEY HAD SPOKEN TO THE CUSTOMER WAS ON (B)(6) 2016. MULTIPLE UNSUCCESSFUL ATTEMPTS WERE MADE TO CONTACT THE NEXT OF KIN TO OBTAIN FURTHER INFORMATION. A CALL BACK REQUEST LETTER HAS BEEN SENT ON (B)(6) 2016. NO CALL BACK HAS BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 723616 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723RNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Death |