FDA Adverse Event Malfunction Summary report: N

SPINNING SPIROS

MDR report key: 6072231 · Received November 2, 2016

Report

Report Number
2025816-2016-00211
Event Type
Malfunction
Date Received
November 2, 2016
Date of Event
September 23, 2016
Report Date
November 2, 2016
Manufacturer
ICU MEDICAL, INC.
Product Code
FPA
PMA / PMN Number
K082806
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LOT REVIEW: LOT# 324410 SHOWED (B)(4) UNITS WERE MANUFACTURED, TESTED, INSPECTED AND RELEASED IN APRIL 2016, THERE WERE NO EXCEPTION DOCUMENTS. VISUAL RECEIPT: ON 10/11/2016 RECEIVED THE FOLLOWING: ONE USED CH2000S, SPINNING SPIROS, LOT# 2844003, ONE USED 100ML IV ADMIN CONTAINER, ONE USED ADMIN SET. THE SPIROS WAS NOT CONNECTED TO ANOTHER DEVICE. THE SPIN FEATURE OF THE SPIROS HAD BEEN ACTIVATED. FUNCTIONAL TESTING: A SINGLE USED CH2000S SPINNING SPIROS WAS RETURNED SEPARATE FROM A USED ADMINISTRATION SET. THE SPIN FEATURE WAS ACTIVATED AND SPINNING FREELY. THE CH2000S, SPINNING SPIROS WAS CONNECTED TO THE MALE LUER LOCK OF THE USED ADMINISTRATION SET AT A CONNECTION TORQUE OF 1.3IN-LBS. THE RESIDUAL DISCONNECT TORQUE WAS MEASURED AT 0.11IN-LBS. THE DISCONNECT TORQUE WAS MEASURED AT 1.01IN-LBS. FINAL ANALYSIS SUMMARY: THE CH2000S SPINNING SPIROS WAS RETURNED DISCONNECTED FROM THE RETURNED ADMINISTRATION SET. WHEN THE CH2000S SPINNING SPIROS WAS RECONNECTED TO THE RETURNED ADMINISTRATION SET THE SPIROS DID NOT SHOW ANY PROPENSITY TO DISCONNECT PREMATURELY.

Description of Event or Problem · 1

COMPLAINT RECEIVED REGARDING ONE, CH2000S, SPINNING SPIROS, LOT# 2844003 (MFD. APRIL 2014). REPORT STATES, SPIROS WAS DISCONNECTED FROM AN INFUSION SET AFTER SPIROS WAS TURNED TO REVERSE. PROPER PPE WAS UTILIZED. NO ADVERSE CLINICIAN OR PATIENT CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
724721 SPINNING SPIROS SPINNING SPIROS FPA ICU MEDICAL, INC. CH2000S 3244010

Patients

Seq Age Sex Outcome Treatment
1