FDA Adverse Event Malfunction Summary report: N

AIRFLOW

MDR report key: 6072180 · Received November 2, 2016

Report

Report Number
2246980-2016-00063
Event Type
Malfunction
Date Received
November 2, 2016
Date of Event
October 4, 2016
Report Date
October 5, 2016
Manufacturer
VENTLAB LLC.
Product Code
BTM
PMA / PMN Number
K012842
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AN EXTENSIVE AND THOROUGH INVESTIGATION WAS PERFORMED BY THE ENGINEERING TEAM THAT CONCLUDED THE ROOT CAUSE FOR THIS PROBLEM IS THE FACT THAT THE PATIENT PORT AND THE MASK ARE MADE OUT OF SIMILAR BASE MATERIALS RESULTING IN AN UNDESIRABLY HIGH COEFFICIENT OF FRICTION BETWEEN THE TWO PARTS. STEPS HAVE BEEN TAKEN TO IMPLEMENT A MATERIAL COMBINATION WITH A LOWER COEFFICIENT OF FRICTION THAT SIGNIFICANTLY REDUCES THE EFFORT REQUIRED TO REMOVE THE MASK FROM THE PATIENT VALVE.

Description of Event or Problem · 1

THE CUSTOMER ALLEGES "THE MASK CANNOT BE REMOVED FROM THE BAG." NO OTHER DETAILS WERE PROVIDED AND NO PATIENT INJURY/HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
722878 AIRFLOW RESUS BAG BTM VENTLAB LLC. AF1140MB-P 304948

Patients

Seq Age Sex Outcome Treatment
1