ASR UNI FEMORAL IMPL SIZE 46
Report
- Report Number
- 1818910-2016-30528
- Event Type
- Injury
- Date Received
- November 1, 2016
- Date of Event
- October 26, 2016
- Report Date
- October 26, 2016
- Manufacturer
- DEPUY INTERNATIONAL LTD - 8010379
- Product Code
- KWA
- PMA / PMN Number
- K040627
- Removal / Correction Number
- Z-1749/1816-2011
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H6 PATIENT CODE: NO CODE AVAILABLE (3191) USED TO CAPTURE THE PATIENT CODE BLOOD HEAVY METAL INCREASED.
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THIS HIP REPLACEMENT PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET AND THE PRODUCT CODES ARE NOW CONSIDERED INACTIVE. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY : NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THIS HIP REPLACEMENT PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET AND THE PRODUCT CODES ARE NOW CONSIDERED INACTIVE. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT : THE PRODUCT INVESTIGATION FOUND NO EVIDENCE SUSPECTING AN ERROR IN THE MANUFACTURING OR MATERIAL THAT WOULD BE A CONTRIBUTING FACTOR IN THE REPORTED ALLEGATION(S). A MANUFACTURING RECORDS EVALUATION (MRE) WAS NOT PERFORMED.
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: UDI (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. CORRECTED: G1.
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. THIS HIP REPLACEMENT PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET AND THE PRODUCT CODES ARE NOW CONSIDERED INACTIVE. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT: PREVIOUS INVESTIGATIONS THAT HAVE INCLUDED MANUFACTURING RECORD EVALUATIONS (MRE) SINCE THE ASR PLATFORM WAS LAUNCHED HAVE SHOWN NO INDICATION OF DEVIATIONS OR ANOMALIES WITH REGARD TO MATERIAL, MANUFACTURING OR INSPECTION.
ASR LITIGATION RECORDS RECEIVED ALLEGING INJURY, ECONOMIC LOSS, LOSS OF SERVICES, ELEVATED LEVELS OF CHROMIUM AND COBALT IN HER BLOOD, PAIN, SUFFERING AND EMOTIONAL DISTRESS.
THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. REF. (B)(4) DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
PRODUCT COMPLAINT # (B)(4). INITIAL REPORTER OCCUPATION: LAWYER. (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).
REVISION NOTES STATED THAT THERE WAS SIGNIFICANT AMOUNT OF SOFT TISSUE INFLAMMATION WITHIN THE JOINT WHICH WAS REMOVED. THERE WAS LARGE CYSTIC CHANGES WITH SIGNIFICANT AMOUNT OF BONE LOSS IN THE ACETABULUM. CUP, HEAD, AND SLEEVE WERE REMOVED. OPERATIVE FINDINGS STATED THAT A SOFT TISSUE WITH METAL DEBRIS WITH GRAY DISCOLORATION NOTED AND SMALL AMOUNT OF FLUID LARGE AMOUNT OF TISSUE HYPERTROPHY NOTED.
PATIENT WAS REVISED DUE TO PAIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 721134 | ASR UNI FEMORAL IMPL SIZE 46 | HIP FEMORAL HEAD | KWA | DEPUY INTERNATIONAL LTD - 8010379 | 2283732 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Required Intervention | ADAPTER SLEEVES 12/14 +2.| ASR ACETABULAR CUPS 52.| CORAIL2 NON COL HO SIZE 12. |