REFOBACIN BONE CEMENT R 1X40
Report
- Report Number
- 3006946279-2016-00407
- Event Type
- Malfunction
- Date Received
- October 31, 2016
- Report Date
- April 27, 2017
- Manufacturer
- BIOMET FRANCE S.A.R.L.
- Product Code
- MBB
- PMA / PMN Number
- PK150850
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. (B)(4). COMPLAINT SAMPLE WAS EVALUATED AND THE REPORTED EVENT WAS CONFIRMED. EVALUATION OF RETURNED DEVICE FOUND OPENED SUPPLIER INNER SEALING. THE MANUFACTURER INNER SEALING WAS NOT DAMAGED. THE OUTER PACKAGING WAS NOT DAMAGED. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. INVESTIGATION RESULTS CONCLUDED THAT THE REPORTED EVENT WAS DUE TO SUPPLIER PACKAGING ISSUE. CORRECTIVE ACTION HAS BEEN INITIATED TO ADDRESS REPORTED ISSUE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
THIS USER FACILITY IS OUTSIDE OF THE UNITED STATES. THE NECESSARY MANUFACTURING HISTORY TO REVIEW WAS NOT PROVIDED. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY. QUANTITY - 2. ZIMMER BIOMET (B)(4) AND PACKAGING SUPPLIER ARE IN THE PROCESS OF INITIATING MODIFICATIONS TO ADDRESS REPORTED ISSUE. CORRECTIVE ACTION HAS BEEN INITIATED TO ADDRESS THE REPORTED ISSUE. THIS REPORT IS NUMBER 1 OF 2 MDRS FILED FOR THE SAME EVENT (REFERENCE 3006946279-2016-00407 / 00408).
IT WAS REPORTED THAT WHEN BOX OF BONE CEMENT WAS OPENED, THE INNER PACKAGE OF POWDER WAS LEAKING INTO THE OUTER PACKAGING. THERE WAS NO PATIENT INJURY AND NO DELAY IN A PROCEDURE AS A RESULT OF THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 717760 | REFOBACIN BONE CEMENT R 1X40 | BONE CEMENT, ANTIBIOTIC | MBB | BIOMET FRANCE S.A.R.L. | N/A | A541AI0108 |