FDA Adverse Event Malfunction Summary report: N

PRIDE MOBILITY PRODUCTS

MDR report key: 6067700 · Received October 31, 2016

Report

Report Number
2530130-2016-00137
Event Type
Malfunction
Date Received
October 31, 2016
Date of Event
July 1, 2016
Report Date
November 8, 2016
Manufacturer
PRIDE MOBILITY PRODUCTS
Product Code
INI
PMA / PMN Number
K092088
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CONTROLLER AND TRANSAXLE ONLY WERE RETURNED AND EVALUATED. THE SMOKE AND BURNING SMELL EXPERIENCED WAS LIKELY THE RESULT OF A PCB FAILURE CONTAINED WITHIN THE METAL S-DRIVE CONTROLLER ENCLOSURE. IT IS UNKNOWN WHAT CAUSED THE PCB TO FAIL BUT THERE WAS VISUAL EVIDENCE THAT THE RETURNED COMPONENTS WERE STORED IN A MOIST AREA OR EXPOSED TO LIQUID.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN MADE AVAILABLE FOR EVALUATION AT THIS TIME. SHOULD FURTHER INFORMATION OR THE DEVICE BECOME AVAILABLE FOR EVALUATION, A FOLLOW-UP REPORT WILL THEN BE ISSUED.

Description of Event or Problem · 1

UNIT WAS IN STORAGE AS IT WAS BACK FOR REPAIR WHEN IT IS ALLEGED IT WAS FOUND SMOKING AND HAD A BURNING SMELL. THE CONTROLLER WAS OPENED AND APPEARED THAT THE INSIDE WAS BURNED COMPLETELY INCLUDING THE SAFETY RELAY.

Description of Event or Problem · 1

UNIT WAS IN STORAGE AS IT WAS BACK FOR REPAIR WHEN IT IS ALLEGED IT WAS FOUND SMOKING AND HAD A BURNING SMELL. THE CONTROLLER WAS OPENED AND APPEARED THAT THE INSIDE WAS BURNED COMPLETELY INCLUDING THE SAFETY RELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
718223 PRIDE MOBILITY PRODUCTS MOTORIZED THREE WHEELED VEHICLE INI PRIDE MOBILITY PRODUCTS LUNA VICTORY ECONOMY S613E N/A

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening