PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2016-38682
- Event Type
- Malfunction
- Date Received
- October 31, 2016
- Date of Event
- October 9, 2016
- Report Date
- October 9, 2016
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
INSULIN PUMP PASSED OPERATING CURRENT AND DISPLACEMENT TEST HOWEVER COMPROMISED FORCE SENSOR SYSTEM ALARM DURING PRIME TEST AT 4 POUNDS AND MOTOR ERROR ALARM DURING THE BASIC OCCLUSION TEST DUE TO LOOSE/PROTRUDED DRIVE SUPPORT DISK. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. INSULIN PUMP RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED LCD WINDOW, BROKEN BELT CLIP SLOT AND BROKEN BATTERY TUBE THREADS. (B)(4).
CUSTOMER REPORTED VIA PHONE CALL THAT THE DEVICE ALARMED COMPROMISED FORCE SENSOR SYSTEM ALARM. CUSTOMER'S BLOOD GLUCOSE WAS 232 MG/DL. DRIVE SUPPORT CAP WAS STICKING OUT. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROFESSIONALS' INSTRUCTIONS. CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WILL BE REPLACED. CUSTOMER AGREED TO RETURN THE DEVICE FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 717166 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |