O-ARM 1000 IMAGING SYSTEM
Report
- Report Number
- 1723170-2016-03218
- Event Type
- Malfunction
- Date Received
- October 28, 2016
- Date of Event
- June 25, 2015
- Report Date
- October 28, 2016
- Manufacturer
- MEDTRONIC NAVIGATION, INC. (LITTLETON)
- Product Code
- OXO
- PMA / PMN Number
- K050996
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
NO PATIENT INFORMATION PROVIDED AS NO PATIENT WAS INVOLVED IN THIS CONCERN. SITE'S BIOMEDICAL ENGINEER REPORTED THAT THE SYSTEM DISPLAYED 'FRAME RATE ERROR' AFTER SYSTEM BOOT UP. THE IMAGE ACQUISITION SYSTEM (IAS) AND THE MOBILE VIEW STATION (MVS) WERE RESTATED AND THE UMBILICAL CABLE WAS RECONNECTED WITHOUT RESOLUTION. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED. THE MOBILE VIEW STATION (MVS) UMBILICAL CABLE WAS RETURNED TO MEDTRONIC FOR ANALYSIS. FOLLOWING ANALYSIS RESULTS, IT WAS FOUND THAT THE SCREWS ON THE PLEORA END ARE MISSING, CAUSING A LOOSE INTERMITTENT CONNECTION. THE CABLE WAS REPLACED AND THE ISSUE WAS RESOLVED. A FULL IMAGING SYSTEM CHECK-OUT WAS COMPLETED AND ALL TESTS PASSED. FULL SYSTEM FUNCTIONALITY WAS CONFIRMED AND THE SYSTEM WAS RETURNED TO SERVICE. THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW AS DISCUSSED WITH THE FDA (B)(6) DISTRICT OFFICE ON APRIL 7, 2016 VIA MEDTRONIC NAVIGATION, INC. RESPONSE TO 3/17/2016 (B)(4). THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE REPORTED MALFUNCTION WAS NOT PREVIOUSLY ASSOCIATED WITH SERIOUS INJURIES. SIMILAR MALFUNCTIONS THAT HAD RESULTED IN SIGNIFICANT DELAYS OF SURGERY OR REQUIRED ADDITIONAL INTERVENTION HAD BEEN REPORTED AS MDR SERIOUS INJURIES. BASED ON THIS 483 OBSERVATION, THE REPORTING DETERMINATION HAS BEEN REVISED TO CONSIDER THESE MALFUNCTION EVENTS REPORTABLE. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE APRIL 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.
SITE'S BIOMEDICAL ENGINEER REPORTED THAT THE SYSTEM DISPLAYED 'FRAME RATE ERROR' AFTER SYSTEM BOOT UP. THE IMAGE ACQUISITION SYSTEM (IAS) AND THE MOBILE VIEW STATION (MVS) WERE RESTATED AND THE UMBILICAL CABLE WAS RECONNECTED WITHOUT RESOLUTION. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 715800 | O-ARM 1000 IMAGING SYSTEM | IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE | OXO | MEDTRONIC NAVIGATION, INC. (LITTLETON) | BI-700-00027-120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |