FDA Adverse Event Malfunction Summary report: N

E SERIES DEFIBRILLATOR

MDR report key: 6063260 · Received October 28, 2016

Report

Report Number
1220908-2016-02678
Event Type
Malfunction
Date Received
October 28, 2016
Report Date
October 11, 2016
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K042007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY FILTER-INDUCTOR ON THE SYSTEM BOARD. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
716377 E SERIES DEFIBRILLATOR E SERIES MKJ ZOLL MEDICAL CORPORATION E SERIES NA

Patients

Seq Age Sex Outcome Treatment
1