FDA Adverse Event Malfunction Summary report: N

DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 6062330 · Received October 27, 2016

Report

Report Number
3004753838-2016-09191
Event Type
Malfunction
Date Received
October 27, 2016
Date of Event
October 7, 2016
Report Date
October 8, 2016
Manufacturer
DEXCOM, INC.
Product Code
MDS
UDI-DI
00386270000262
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON (B)(6) 2016, THAT ON (B)(6) 2016, THE RECEIVER DISPLAYED ERR68. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND "CALL TECH SUPPORT" ERROR WAS OBSERVED ON THE SCREEN. THE DATA LOG COULD NOT BE RETRIEVED AND FUNCTIONAL TESTING COULD NOT BE PERFORMED BECAUSE OF THE "CALL TECH SUPPORT" ERROR. THE CUSTOMER COMPLAINT WAS CONFIRMED BECAUSE ERROR ICON WAS DISPLAYED. THE ROOT CAUSE COULD NOT BE DETERMINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
712515 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR MDS DEXCOM, INC. MT22719-PNK 5215855 00386270000262

Patients

Seq Age Sex Outcome Treatment
1 8 YR