FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 6057768 · Received October 26, 2016

Report

Report Number
1000113657-2016-01780
Event Type
Malfunction
Date Received
October 26, 2016
Date of Event
September 29, 2016
Report Date
November 14, 2016
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PRODUCT DOES NOT RETURN FOR EVALUATION YET. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.

Additional Manufacturer Narrative · 1

(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED: THE RETURNED METER AND TEST STRIPS DID MEET ALL THE TESTING PERFORMANCE SPECIFICATIONS.THE PRODUCT SYSTEM IS FUNCTIONING PROPERLY AS INTENDED. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION:USER HAD AN INACCURATE REFERENCE.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 203, 211 AND 267 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 110-130 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 218 AND 187 MG/DL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 8/23/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 203, 211 AND 267 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 110-130 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 218 AND 187 MG/DL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 8/23/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: 203MG/DL (B)(6) 2016 09:20 AM FASTING: YES; 182MG/DL (B)(6) 2016 06:48 AM FASTING: YES; 208MG/DL (B)(6) 2016 11:10 PM FASTING: YES; 211MG/DL (B)(6) 2016 06:54 PM FASTING: YES; 267MG/DL (B)(6) 2016 04:08 PM FASTING: YES. MEMORY CONCERNS: ALL RESULTS LISTED ABOVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
708366 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MT1635

Patients

Seq Age Sex Outcome Treatment
1 0 YR SECOND THERAPY