TRUEMETRIX
Report
- Report Number
- 1000113657-2016-01780
- Event Type
- Malfunction
- Date Received
- October 26, 2016
- Date of Event
- September 29, 2016
- Report Date
- November 14, 2016
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI
- Reporter Occupation
- OTHER
Narratives
(B)(4). PRODUCT DOES NOT RETURN FOR EVALUATION YET. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED: THE RETURNED METER AND TEST STRIPS DID MEET ALL THE TESTING PERFORMANCE SPECIFICATIONS.THE PRODUCT SYSTEM IS FUNCTIONING PROPERLY AS INTENDED. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION:USER HAD AN INACCURATE REFERENCE.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 203, 211 AND 267 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 110-130 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 218 AND 187 MG/DL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 8/23/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 203, 211 AND 267 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 110-130 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 218 AND 187 MG/DL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 8/23/2017 AND OPEN VIAL DATE IS (B)(6) 2016. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: 203MG/DL (B)(6) 2016 09:20 AM FASTING: YES; 182MG/DL (B)(6) 2016 06:48 AM FASTING: YES; 208MG/DL (B)(6) 2016 11:10 PM FASTING: YES; 211MG/DL (B)(6) 2016 06:54 PM FASTING: YES; 267MG/DL (B)(6) 2016 04:08 PM FASTING: YES. MEMORY CONCERNS: ALL RESULTS LISTED ABOVE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 708366 | TRUEMETRIX | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | TRUEMETRIX | MT1635 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR | SECOND THERAPY |