FDA Adverse Event Malfunction Summary report: N

INDIGO SYSTEM SEPARATOR 8

MDR report key: 6056162 · Received October 25, 2016

Doing adverse event research by hand?

The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.

Free to try · every answer cites its records

Report

Report Number
3005168196-2016-01513
Event Type
Malfunction
Date Received
October 25, 2016
Date of Event
September 25, 2016
Report Date
September 26, 2016
Manufacturer
PENUMBRA, INC.
Product Code
DXE
UDI-DI
00814548016153
PMA / PMN Number
K161523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS AVAILABLE FOR RETURN. A FOLLOW UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE INVESTIGATION.

Additional Manufacturer Narrative · 1

PMA/510(K)# HAS BEEN UPDATED TO REFLECT THE MOST CURRENT 510(K) # FOR THE DEVICE AT THE TIME OF THE REPORTED COMPLAINT. RESULTS: THE INDIGO SYSTEM SEPARATOR 8 (SEP8) CORE WIRE WAS FRACTURED APPROXIMATELY 148.5 CM FROM THE PROXIMAL END. THE SEP8 BULB WAS STILL INTACT WITH THE COIL WINDINGS OF THE SEP8. THE SEP8 COIL WINDINGS WERE SEPARATED BY THE INVESTIGATOR FOR FURTHER EVALUATION. THE DISTAL SEGMENT OF THE FRACTURED CORE WIRE WAS INTACT WITH THE DISTAL TIP OF THE BULB TO THE DISTAL SOLDER JOINT. CONCLUSIONS: EVALUATION OF THE RETURNED DEVICE REVEALED THAT THE CORE WIRE TO THE SEP8 WAS FRACTURED PROXIMAL TO THE BULB. THIS TYPE OF DAMAGE TYPICALLY OCCURS DUE TO IMPROPER HANDLING DURING USE. IF THE SEP8 IS WITHDRAWN FORCEFULLY AGAINST RESISTANCE, DAMAGE SUCH AS THIS MAY OCCUR. SEP8 DEVICES ARE VISUALLY INSPECTED DURING IN-PROCESS INSPECTION AND DURING QUALITY INSPECTION AFTER MANUFACTURING. THE MANUFACTURING RECORDS FOR THIS LOT WERE REVIEWED AND DID NOT REVEAL ANY OUTSTANDING DISCREPANCIES, DESIGN, OR QUALITY CONCERNS.

Description of Event or Problem · 1

THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE TO TREAT A PULMONARY EMBOLISM (PE) USING AN INDIGO SYSTEM SEPARATOR 8 (SEP8). DURING THE PROCEDURE, THE PHYSICIAN PLACED A NON-PENUMBRA SHEATH IN THE LEFT CORONARY ARTERY AND THEN USED AN INDIGO SYSTEM ASPIRATION CATHETER 8 (CAT8) WITH THE SEP8 TO REMOVE THROMBUS. AFTER USING THE SEP8 WITH THE CAT8 FOR SEVERAL PASSES, THE PHYSICIAN EXPERIENCED RESISTANCE AND SUBSEQUENTLY, THE SEP8 BECAME STUCK IN THE CAT8 AND WOULD NOT MOVE FORWARD OR BACKWARD. THEREFORE, THE CAT8 WITH THE SEP8 STILL INSIDE OF IT WAS REMOVED FROM THE PATIENT'S BODY. WHILE THE DEVICES WERE ON THE BACK TABLE, THE PHYSICIAN REMOVED THE SEP8 FROM THE CAT8 AND NOTICED THAT THE WIRE IN FRONT OF THE SEP8 BULB APPEARED TO BE FRACTURED AND UNWINDING. THE PROCEDURE WAS THEN COMPLETED USING A NEW SEP8 AND THE SAME CAT8. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
705098 INDIGO SYSTEM SEPARATOR 8 DXE DXE PENUMBRA, INC. F66926 00814548016153

Patients

Seq Age Sex Outcome Treatment
1 45 YR