HAKIM IN LINE PROGRAMMABLE VALVE
Report
- Report Number
- 1226348-2016-10746
- Event Type
- Malfunction
- Date Received
- October 25, 2016
- Date of Event
- October 1, 2016
- Manufacturer
- CODMAN & SHURTLEFF, INC. / MEDOS S.A.
- Product Code
- JXG
- PMA / PMN Number
- K974739
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
(B)(4). UPON COMPLETION OF THE INVESTIGATION, A FOLLOW UP REPORT WILL BE FILED.
UPON COMPLETION OF THE INVESTIGATION, IT WAS NOTED THAT THE IMAGES WERE TAKEN OF THE ¿AS RECEIVED¿ VALVE. THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS AT 90MMH2O. THE VALVE WAS HYDRATED FOR 24 HOURS. THE VALVE WAS VISUALLY INSPECTED: A TEAR IN THE SILICONE HOUSING WAS NOTED, AS WELL AS ABRASIONS MARKS IN THE SILICONE HOUSING OVER THE VALVE CASING. THIS IS COULD BE DUE TO UNSHOD FORCEPS BEING USED. NO DEFECTS WERE NOTED WITH THE BACTISEAL CATHETER. THE VALVE WAS FLUSHED, FAILED LEAKED FROM TEAR/CUT IN SILICONE HOUSING. THE VALVE WAS LEAK TESTED, FAILED LEAKED FROM TEAR/CUT IN SILICONE HOUSING. REVIEW OF THE HISTORY DEVICE RECORDS CONFIRMED THE VALVE PRODUCT CODE 82-3162, WITH LOT CVDBNC, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK IN 28TH APRIL 2016. THE ROOT CAUSE TO THE TEAR/CUT IN THE SILICONE HOUSING IS PROBABLY BEING DUE TO THE USER, THIS HOWEVER COULD NOT BE DETERMINED. AS NOTED IN THE IFU SILICONE HAS A LOW TEAR / CUT RESISTANCE. PER HHE, IT HAS BEEN CONCLUDED THAT SILICONE HOUSING TEARS ARE NOT DESIGN RELATED, IN ORDER FOR THE HOUSING TEAR TO OCCUR THE USER HAS TO COMPROMISE THE SILICONE THROUGH A NICK OR TEAR IN ORDER FOR THE EVENT TO OCCUR. VALIDATION TESTING DEMONSTRATES A ROBUST DESIGN. TESTING HAS SHOWN THAT IF THE SILICONE HOUSING HAS BEEN COMPROMISED, A HOUSING TEAR IS LIKELY TO OCCUR. NO DEFECTS WERE NOTED WITH THE BACTISEAL CATHETER. BASED ON THE RESULTS OF THIS INVESTIGATION, NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME, THIS COMPLAINT IS CLOSED.
TODAY WHILE DOING A PROGRAMMABLE SHUNT IN (B)(6) HOSPITAL, THE CLINICIAN FOUND THAT THERE WAS A LEAKAGE FROM THE SIDE OF THE CHAMBER WHICH IS VERY UNUSUAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 705885 | HAKIM IN LINE PROGRAMMABLE VALVE | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF, INC. / MEDOS S.A. | CVDBNC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82-3072 |