PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 6054670 · Received October 25, 2016

Report

Report Number
2032227-2016-34475
Event Type
Injury
Date Received
October 25, 2016
Date of Event
October 19, 2016
Report Date
April 6, 2017
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT ESCAPE, ACT, EXPRESS, UP ARROW AND DOWN ARROW BUTTONS DUE TO FLATTENED DOME SWITCHES. ALL OPERATING CURRENTS WITHIN SPECIFICATIONS AND PASSED FUNCTIONAL TESTING INCLUDING THE REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, EXCESSIVE NO DELIVERY TEST, SELF-TEST, A21 ERROR TEST AND DISPLACEMENT TEST. NO EXCESSIVE NO DELIVERY ALARMS NOTED. THE INSULIN PUMP HAD CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND MINOR SCRATCHED LCD WINDOW.

Additional Manufacturer Narrative · 1

THE PUMP PASSED THE DELIVERY ACCURACY TEST.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE CALL THAT SHE WAS HOSPITALIZED FOR DIABETIC KETOACIDOSIS DUE TO HIGH BLOOD GLUCOSE. CUSTOMER'S BLOOD GLUCOSE WAS 476 MG/DL. CUSTOMER'S BLOOD GLUCOSE AT TIME OF CALL WAS 337 MG/DL. CUSTOMER MENTIONED THE BUTTONS WERE UNRESPONSIVE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROFESSIONALS' INSTRUCTIONS. CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WILL BE REPLACED. CUSTOMER AGREED TO RETURN THE DEVICE FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
705436 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAP

Patients

Seq Age Sex Outcome Treatment
1 23 YR Hospitalization