- Report Number
- 3010673777-2016-00003
- Event Type
- Injury
- Date Received
- October 25, 2016
- Report Date
- October 25, 2016
- Manufacturer
- NOVASSTEP
- Product Code
- JDR
- UDI-DI
- 03700879501291
- PMA / PMN Number
- K142111
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
COMPLAINT # (B)(4) OPENED : ALL INVESTIGATION DONE AT THIS TIME CONFIRMED THESE BATCHES OF ARCAD WERE PRODUCED AND CONTROLLED CONFORM TO NOVASTEP SPECIFICATIONS. COMMENT: 2 STAPLES HAVE BEEN USED FOR THIS CASE, WE DON'T KNOW WHICH STAPLE LOT NUMBER HAS BROKEN. FOR INFORMATION, WE ADD IN THIS MDR ALL INFORMATION RELATED THE 2ND STAPLE. MODEL # CATALOG # : CS041618, LOT # : D00213, EXPIRY DATE : 02/01/2016. DEVICE EVALUATED BY MANUFACTURER ? NOT EXPLANTED. DEVICE NOT EXPLANTED.
DURING A SALES REP VISIT ON (B)(6) 2016 IT WAS REPORTED TO NOVASTEP SALES REP. THAT DURING THE LAST 12 MONTH 3 CASES HAVE CONDUCTED TO ARCAD 18 COMPRESSIVE STAPLE BREAKAGE. THIS IS THE SAME SURGEON FOR THESE 3 CASES. FIRST CASE: ARCAD 18-16-18 COMPRESSIVE STAPLE WAS DETECTED AS BROKEN BY THE SURGEON DURING THE ANALYSIS OF XRAY PERFORMED DURING A PATIENT FOLLOW UP VISIT. THE SURGEON EXPECTS THAT THE USE OF ONLY TWO ARCAD 18 STAPLES TO PROCEED MTP ARTHRODESIS TO BE TOO WEAK. ASSEMBLY SHOULD HAVE BEEN DONE USING BIGGER STAPLES OR ADDITIONAL DEVICES ASSOCIATED WITH TWO ARCAD 18 COMPRESSIVE STAPLES. THE OTHERS 2 CASES HAVE BEEN REPORTED IN REPORT 3010673777-2016-00001, 3010673777-2016-00002.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 705724 | ARCAD | STAPLE | JDR | NOVASSTEP | ARCAD | C00684 | 03700879501291 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |