ANIMAS VIBE
Report
- Report Number
- 2531779-2016-29592
- Event Type
- Malfunction
- Date Received
- October 25, 2016
- Report Date
- October 2, 2016
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- OYC
- PMA / PMN Number
- P130007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/07/2016 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE BATTERY COMPARTMENT/CAP ARE UNDAMAGED AND ABLE TO FIT SECURELY. THE PUMP POWERS UP WITH AUDITORY AND VIBRATION TO A BLANK SCREEN, UNABLE TO VERIFY STEPS AND TO ADEQUATELY INVESTIGATE REPORTED ¿POWER¿ COMPLAINT. THE PUMP¿S COVER WAS REMOVED, U22 EEPROM COMPONENT WAS FOUND CRACKED. UNITY TEST CODE DOWNLOADER TOOL APPLICATION V 1.0.0 WAS USED TO UPLOAD TEST CODE V 1.0.7 TO MASTER/SLAVE/PERIPH PROCESSORS AS PER (6001112). THE UPLOAD WAS SUCCESSFUL, THE PUMP POWERS UP AND DISPLAY JUST ¿MSP¿ INSTEAD OF ¿MSP2¿. (2) IS THE EEPROM COMMUNICATING PROPERLY HENCE THE DIAGNOSIS THAT THE FAILURE IS IN THIS AREA SINCE IT IS MISSING FROM THE DISPLAY. EEPROM FAILURE IS CONCLUDED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 704210 | ANIMAS VIBE | INSULIN INFUSION PUMP | OYC | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR |