FDA Adverse Event
Malfunction
Summary report: N
HEARTSINE SAMARITAN 300P AND PAD PAK
MDR report key: 6052997
·
Received October 25, 2016
Report
- Report Number
- 3004123209-2016-01523
- Event Type
- Malfunction
- Date Received
- October 25, 2016
- Date of Event
- September 29, 2016
- Report Date
- November 29, 2016
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- PMA / PMN Number
- K041067
- Removal / Correction Number
- Z-0124-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 0
EXEMPTION NUMBER E2015058 HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER). H3 OTHER TEXT : NOT YET RETURNED TO MANUFACTURER
Additional Manufacturer Narrative · 0
EXEMPTION NUMBER E2015058. ROUTINE TESTING CONFIRMED THAT THE PAD-PAK WAS FIRST WAS INSTALLED BY THE CUSTOMER ON (B)(6) 2010. THE DEVICE RECORDS MANUAL POWER UPS OF TEN MINUTES DURATION BETWEEN (B)(6) 2016. INVESTIGATION FOUND THE FAULT CAN BE ATTRIBUTED TO MEMBRANE FAILURE. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
Description of Event or Problem · 0
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. DEVICE SWITCHES ON AUTOMATICALLY.