FDA Adverse Event Malfunction Summary report: N

ACCESSORIES FOR SPACE PUMPS

MDR report key: 6052388 · Received October 24, 2016

Report

Report Number
9610825-2016-00677
Event Type
Malfunction
Date Received
October 24, 2016
Report Date
September 29, 2016
Manufacturer
B. BRAUN MELSUNGEN AG
Product Code
MRZ
PMA / PMN Number
K083689
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2011009. B. BRAUN MEDICAL INC. IS SUBMITTING A SINGLE REPORT ON BEHALF OF B. BRAUN MELSUNGEN (THE MANUFACTURER), AND B. BRAUN MEDICAL INC. (THE IMPORTER). THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). THE PROVIDED PICTURES WERE EXAMINED. ACCORDING TO THE PICTURES THE POWER SUPPLY SEEMED TO BE IN A BAD CONDITION AND SHOWED HEAVY MARKS OF USE. THE MALE CONNECTOR OF THE EXTENSION CABLE WAS MELTED DOWN. ASSUMEDLY AN HIGH AMOUNT OF FLUID CAUSED THIS THERMAL DAMAGE. FLUID HAS TO BE INGRESSED BETWEEN THE MALE CONNECTOR OF THE EXTENSION CABLE AND THE FEMALE CONNECTOR OF THE POWER SUPPLY UNIT. BASED ON FORMER TESTS, THIS ERROR PATTERN IS A RESULT OF USAGE OUT OF THE POWER SUPPLIES SPECIFICATION. IF ADDITIONAL PERTINENT INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2011009. B. BRAUN MEDICAL INC. IS SUBMITTING A SINGLE REPORT ON BEHALF OF B. BRAUN MELSUNGEN (THE MANUFACTURER), AND B. BRAUN MEDICAL INC. (THE IMPORTER). THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). THE DEVICE INVOLVED HAS NOT BEEN RECEIVED FOR EVALUATION AND THE INVESTIGATION IS ONGOING AT THIS TIME. A FOLLOW UP WILL BE SUBMITTED WHEN THE INVESTIGATION RESULTS BECOME AVAILABLE.

Description of Event or Problem · 0

AS REPORTED BY THE USER FACILITY: EVENT 3: THE CUSTOMER IS REPORTING THERMAL DAMAGE. NO INJURIES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
701970 ACCESSORIES FOR SPACE PUMPS US POWER CORD INCL. EXTENSION MRZ B. BRAUN MELSUNGEN AG

Patients

Seq Age Sex Outcome Treatment
1