ANIMAS VIBE
Report
- Report Number
- 2531779-2016-29557
- Event Type
- Malfunction
- Date Received
- October 24, 2016
- Report Date
- October 5, 2016
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- MDS
- PMA / PMN Number
- P130007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/17/2017 WITH THE FOLLOWING FINDINGS: ON INVESTIGATION, THE BATTERY COMPARTMENT WAS CONFIRMED TO BE CRACKED. ON INVESTIGATION, THE PUMP POWERED ON WITH A DIM/FADED/DISCOLORED DISPLAY SCREEN. INVESTIGATION CONFIRMED THE COMPLAINT.
ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 702862 | ANIMAS VIBE | INSULIN INFUSION PUMP | MDS | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |