FDA Adverse Event Malfunction Summary report: N

V60 VENTILATOR

MDR report key: 6051581 · Received October 24, 2016

Report

Report Number
2031642-2016-02849
Event Type
Malfunction
Date Received
October 24, 2016
Report Date
September 28, 2016
Manufacturer
RESPIRONICS CALIFORNIA, INC
Product Code
MNT
PMA / PMN Number
K082660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
RESPIRATORY THERAPIST

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT WAS (B)(6) 2016.

Additional Manufacturer Narrative · 1

THE POWER MANAGEMENT BOARD WAS RETURNED TO THE MANUFACTURER FOR FAILURE INVESTIGATION. VISUAL INSPECTION OF POWER MANAGEMENT (PM) PCBA REVEALED EVIDENCE OF C258 BURN DAMAGE AND TRACES BURN DAMAGE. FAILURE INVESTIGATION DETERMINED THAT THE C258 BURN DAMAGE AND C258 TRACE TO U50 PIN1 OPEN DAMAGE ON THE POWER MANAGEMENT BOARD PREVENTED THE VENTILATOR FROM POWERING ON.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
701883 V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE MNT RESPIRONICS CALIFORNIA, INC V60

Patients

Seq Age Sex Outcome Treatment
1