FDA Adverse Event
Malfunction
Summary report: N
V60 VENTILATOR
MDR report key: 6051581
·
Received October 24, 2016
Report
- Report Number
- 2031642-2016-02849
- Event Type
- Malfunction
- Date Received
- October 24, 2016
- Report Date
- September 28, 2016
- Manufacturer
- RESPIRONICS CALIFORNIA, INC
- Product Code
- MNT
- PMA / PMN Number
- K082660
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- RESPIRATORY THERAPIST
Narratives
Additional Manufacturer Narrative · 1
DATE OF EVENT WAS (B)(6) 2016.
Additional Manufacturer Narrative · 1
THE POWER MANAGEMENT BOARD WAS RETURNED TO THE MANUFACTURER FOR FAILURE INVESTIGATION. VISUAL INSPECTION OF POWER MANAGEMENT (PM) PCBA REVEALED EVIDENCE OF C258 BURN DAMAGE AND TRACES BURN DAMAGE. FAILURE INVESTIGATION DETERMINED THAT THE C258 BURN DAMAGE AND C258 TRACE TO U50 PIN1 OPEN DAMAGE ON THE POWER MANAGEMENT BOARD PREVENTED THE VENTILATOR FROM POWERING ON.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE DEVICE WOULD NOT POWER ON. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 701883 | V60 VENTILATOR | VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE | MNT | RESPIRONICS CALIFORNIA, INC | V60 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |