FDA Adverse Event Malfunction Summary report: N

HF-CABLE, MONOPOLAR, 4 M, UES-30/-40,

MDR report key: 6050029 · Received October 24, 2016

Report

Report Number
2951238-2016-00812
Event Type
Malfunction
Date Received
October 24, 2016
Date of Event
September 30, 2016
Report Date
October 23, 2016
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
GCJ
PMA / PMN Number
PK944201
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED TO OLYMPUS FOR EVALUATION; HOWEVER, A PHOTOGRAPHIC IMAGE OF THE CORD WAS PROVIDED BY THE USER FACILITY. THE IMAGE ILLUSTRATED THAT CORD WAS SEPARATED FROM THE CONNECTOR. OLYMPUS FOLLOWED UP WITH THE USER FACILITY VIA TELEPHONE AND IN WRITING TO OBTAIN ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT BUT WITH NO RESULT. THE MOST LIKELY CAUSE FOR THE REPORTED EVENT IS DUE TO USER HANDLING. THE INSTRUCTION MANUAL GIVES SEVERAL WARNING TO PREVENT THIS TYPE OF CORD DAMAGE. ¿DO NOT USE A CABLE WITH BRITTLE OR DEFECTIVE INSULATION. REPLACE THE CABLE. VISUALLY INSPECT THE CABLE AND THE PLUGS FOR IRREGULARITIES ON THE SURFACE. WHEN USED AS INTENDED THIS PRODUCT IS MORE OR LESS SUBJECT TO WEAR DEPENDING ON THE INTENSITY OF USE. DO NOT USE THE HF CABLE AFTER ONE YEAR OF USE. IN CASE A PRODUCT HAS EXTERNALLY VISIBLE DEFECTS CONTACT THE RESPONSIBLE OLYMPUS REPRESENTATIVE AT ONCE.¿

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT DURING AN UNSPECIFIED PROCEDURE, THE HF-CABLE AND THE DRAPE ON THE PATIENT CAUGHT FIRE. IT WAS REPORTED THAT THE FIRE STARTED BEHIND THE BANANA PLUG (CONNECTION POINT) WHILE IN USE WITH A NON-OLYMPUS GENERATOR. THE FIRE WAS DISTINGUISHED WITH NO USER OR PATIENT INJURY. IT IS UNKNOWN IF THE INTENDED PROCEDURE WAS COMPLETED. ADDITIONALLY, THE USER FACILITY REPORTED THAT HF-CABLE WAS INSPECTED PRIOR TO THE PROCEDURE AND NO ANOMALIES WERE FOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
701634 HF-CABLE, MONOPOLAR, 4 M, UES-30/-40, HF-CABLE GCJ OLYMPUS WINTER & IBE GMBH A0393 307574

Patients

Seq Age Sex Outcome Treatment
1 VALLEY LAB FORCE FX GENERATOR