FDA Adverse Event Malfunction Summary report: N

LAP-BAND AP SYSTEM SMALL

MDR report key: 6049110 · Received October 21, 2016

Report

Report Number
3006722112-2016-00308
Event Type
Malfunction
Date Received
October 21, 2016
Date of Event
September 14, 2016
Report Date
September 23, 2016
Manufacturer
APOLLO ENDOSURGERY, INC.
Product Code
LTI
PMA / PMN Number
P000008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

TAPER II. MEDWATCH SENT TO THE FDA ON 10/21/2016. DEVICE EVALUATION SUMMARY: THE DEVICE WAS RETURNED TO APOLLO FOR ANALYSIS, AND VISUAL EXAMINATION CONFIRMED THE CONNECTOR TYPE AS TAPER II. BROWN DISCOLORATION WAS OBSERVED ON THE ACCESS PORT. A PORT LEAK TEST WAS PERFORMED AND NO LEAKAGE WAS OBSERVED. A FILL INSPECTION TEST WAS PERFORMED AND NO BLOCKAGE OR RESISTANCE TO FLOW WAS OBSERVED. MICROSCOPIC ANALYSIS NOTED A NON-PENETRATING NICK ON THE BAND TUBING. THE BAND TUBING HAD STRIATIONS, CONSISTENT WITH SURGICAL END CUTS TO REMOVE THE DEVICE. DEVICE LABELING ADDRESSES THE REPORTED EVENT OF LEAK AS FOLLOWS: PRECAUTIONS: CARE MUST BE TAKEN DURING BAND ADJUSTMENT TO AVOID PUNCTURING THE TUBING THAT CONNECTS THE ACCESS PORT AND BAND, AS THIS WILL CAUSE LEAKAGE AND DEFLATION OF THE INFLATABLE SECTION. FAILURE TO CREATE A STABLE, SMOOTH PATH FOR THE ACCESS PORT TUBING, WITHOUT SHARP TURNS OR BENDS, CAN RESULT IN TUBING BREAKS AND LEAKAGE. IN ORDER TO AVOID INCORRECT PLACEMENT, THE PORT SHOULD BE PLACED LATERAL TO THE TROCAR OPENING. A POCKET MUST BE CREATED FOR THE PORT SO THAT IT IS PLACED FAR ENOUGH FROM THE TROCAR PATH TO AVOID ABRUPT KINKING OF THE TUBING. THE TUBING PATH SHOULD POINT IN THE DIRECTION OF THE ACCESS PORT CONNECTOR SO THAT THE TUBING WILL FORM A STRAIGHT LINE WITH A GENTLE ARCHING TRANSITION INTO THE ABDOMEN. ADVERSE EVENTS: IT IS IMPORTANT TO DISCUSS ALL POSSIBLE COMPLICATIONS AND ADVERSE EVENTS WITH YOUR PATIENT. COMPLICATIONS WHICH MAY RESULT FROM THE USE OF THIS PRODUCT INCLUDE THE RISKS ASSOCIATED WITH THE MEDICATIONS AND METHODS UTILIZED IN THE SURGICAL PROCEDURE, THE RISKS ASSOCIATED WITH ANY SURGICAL PROCEDURE AND THE PATIENT'S DEGREE OF INTOLERANCE TO ANY FOREIGN OBJECT IMPLANTED IN THE BODY. DEFLATION OF THE BAND MAY OCCUR DUE TO LEAKAGE FROM THE BAND, THE PORT OR THE CONNECTING TUBING. WARNINGS: PATIENTS SHOULD BE ADVISED THAT THE LAP-BAND AP® SYSTEM IS A LONG-TERM IMPLANT. EXPLANT AND REPLACEMENT SURGERY MAY BE INDICATED AT ANY TIME. MEDICAL MANAGEMENT OF ADVERSE REACTIONS MAY INCLUDE EXPLANTATION. REVISION SURGERY FOR EXPLANTATION AND REPLACEMENT MAY ALSO BE INDICATED TO ACHIEVE PATIENT SATISFACTION.

Description of Event or Problem · 1

REPORTED AS: A PATIENT WITH THE LAP-BAND SYSTEM WAS REPORTED TO HAVE A "PORT LEAK." THE DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
700327 LAP-BAND AP SYSTEM SMALL ADJUSTABLE GASTRIC BAND LTI APOLLO ENDOSURGERY, INC. AB-20260 2696960

Patients

Seq Age Sex Outcome Treatment
1 49 YR