METRIX AIR
Report
- Report Number
- 1000113657-2016-01763
- Event Type
- Malfunction
- Date Received
- October 21, 2016
- Date of Event
- September 29, 2016
- Report Date
- October 21, 2016
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ
- Reporter Occupation
- OTHER
Narratives
(B)(4). INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECT FOUND ON RETURNED METER. MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION: USER HAD AN INACCURATE REFERENCE.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS FROM RESULTS OBTAINED OF 202, 346, 173 AND 174 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 128-160 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2016, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULTS OF 202 AND 346 MG/DL. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 6/30/2017. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (DATE / TIME NOT SET): 202MG/DL, (B)(6) 2016 07:26 AM, FASTING: YES; 173MG/DL, (B)(6) 2016 07:42 AM, FASTING: YES; 174MG/DL, (B)(6) 2016 07:29 AM, FASTING: YES; 155MG/DL, (B)(6) 2016 06:34 AM, FASTING: YES; 146MG/DL, (B)(6) 2016 07:47 AM, FASTING: YES. MEMORY CONCERNS: 202. THE CUSTOMER CALLED AND STATES THAT HE JUST GOT A HIGHER READING THAN HIS NORMAL RANGE OF 128-160MG/DL ON HIS METER. THE CUSTOMER STATES HE IS FEELING GOOD AND DENIES ANY DIABETIC SYMPTOMS. THE CUSTOMER ALSO STATES THE STRIPS ARE ABSORBING THE BLOOD SAMPLES SLOWER THAN USUAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 698687 | METRIX AIR | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | METRIX AIR | MT1766 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR | SECOND THERAPY |