QC+F/PINS Ø1.5-4
Report
- Report Number
- 8030965-2016-15278
- Event Type
- Malfunction
- Date Received
- October 21, 2016
- Date of Event
- October 7, 2016
- Report Date
- October 7, 2016
- Manufacturer
- SYNTHES OBERDORF
- Product Code
- GEY
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL NARRATIVE: (B)(4). (B)(6). AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND FOUND THAT THE COUPLING TOOL SIDE WAS WORN OUT. IT WAS ALSO NOTED THAT THE COUPLING TOOL WAS SLIGHTLY DIRTY AND THEREFORE, THE SPIKE WIRES WERE NOT CLAMPING PROPERLY. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO PREMATURE WEAR. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
IT WAS REPORTED FROM (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE QUICK COUPLING DEVICE HANDLE WAS STICKING AND WAS NOT GRIPPING THE WIRE. IT WAS REPORTED THAT THERE WAS A FOUR MINUTE DELAY IN THE PROCEDURE DUE TO THE EVENT AND AN UNSPECIFIED SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO PATIENT OR USER INJURIES REPORTED. IT WAS REPORTED THERE WAS NO MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 697702 | QC+F/PINS Ø1.5-4 | MOTOR, SURGICAL INSTRUMENT, AC-POWERED | GEY | SYNTHES OBERDORF | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | WIRE DEVICE |