FDA Adverse Event Injury Summary report: N

ANIMAS VIBE

MDR report key: 6046263 · Received October 21, 2016

Report

Report Number
2531779-2016-29328
Event Type
Injury
Date Received
October 21, 2016
Report Date
October 16, 2016
Manufacturer
ANIMAS CORPORATION
Product Code
OYC
UDI-DI
10840406100051
PMA / PMN Number
P130007
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 2/3/2017. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 01/10/2017 WITH THE FOLLOWING FINDINGS: BLACK BOX SHOWS MULTIPLE CALL SERVICE 078-0008 ALARMS FOLLOWING A REPLACE BATTERY AT 5:26PM 10/16/16 INSULIN DELIVERY IS NOT RESUMED AFTER THIS TIME. DAILY INSULIN DELIVERY TOTALS PRIOR TO THIS TIME CORRECTLY REFLECT USER'S PROGRAMMED BASAL RATES. DURING REWIND PUMP GIVES A CALL SERVICE 078-0008 ALARM. MOISTURE VISIBLE IN BATTERY COMPARTMENT AND DISPLAY. BATTERY COMPARTMENT IS CRACKED ABOVE AND BELOW THE BUMPER PAD. PUMP LEAKS AT BATTERY COMPARTMENT. PUMP OPENED AND MOISTURE DAMAGE FOUND ON PCB. RECURRING ALARM DUE TO MOISTURE DAMAGE.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS AND ALLEGED THAT THE PATIENT HAD A CS 078 ALARM ISSUE, AND THE PATIENT HAD A BLOOD GLUCOSE (BG) OVER 500 MG/DL ,WITH NAUSEA AND VOMITING. THE PATIENT DISCONTINUED PUMP THERAPY INSULIN INJECTIONS WERE GIVEN BY AN HCP. THIS COMPLAINT IS BEING REPORTED AS THE PATIENT EXPERIENCED HYPERGLYCEMIA AND THE ALARM ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
699157 ANIMAS VIBE INSULIN INFUSION PUMP OYC ANIMAS CORPORATION 10840406100051

Patients

Seq Age Sex Outcome Treatment
1 35 YR Life Threatening| R