PRESSURE WIRE AERIS G8
Report
- Report Number
- 3008452825-2016-00142
- Event Type
- Injury
- Date Received
- October 20, 2016
- Date of Event
- September 26, 2016
- Report Date
- September 26, 2016
- Manufacturer
- ST. JUDE MEDICAL, COSTA RICA LTDA
- Product Code
- DXO
- PMA / PMN Number
- K131452
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THE RESULTS OF THE INVESTIGATION CONCLUDED THE COREWIRE WAS FRACTURED INTO TWO SECTIONS AT THE DISTAL TIP. THE TIP COIL HAD BEEN FRACTURED AND SEPARATED FROM THE WELD JOINT. THE PROXIMAL SECTION OF THE COREWIRE HAD BEEN KINKED. THE COMBINED LENGTH OF THE TWO COREWIRE SECTIONS INDICATED A PORTION OF THE COREWIRE WAS NOT RETURNED FROM THE DISTAL TIP ASSEMBLY. THERE WAS NO EVIDENCE FOUND TO SUGGEST THE INCIDENT WAS DUE TO AN INTRINSIC DEFECT IN THE RETURNED DEVICE. THE DEVICE MET SPECIFICATIONS PRIOR TO LEAVING SJM MANUFACTURING FACILITIES AS SUPPORTED BY A REVIEW OF THE DEVICE HISTORY RECORD. THE CAUSE OF THE TIP COIL AND COREWIRE DAMAGE IS CONSISTENT WITH FORCIBLE CONTACT DURING USE. THE PRESSUREWIRE INSTRUCTIONS FOR USE (IFU) STATES THAT EXCESSIVE MANIPULATION OF THE PRESSUREWIRE WHEN THE SENSOR ELEMENT OR PRESSUREWIRE TIP IS LOCATED IN SHARP BEND MAY CAUSE DAMAGE OR TIP FRACTURE. THE PRESSUREWIRE INSTRUCTIONS FOR USE (IFU) STATES THAT TORQUEING THE PRESSUREWIRE AGAINST RESISTANCE OR REPEATED ATTEMPTS TO CROSS A TOTAL VESSEL OCCLUSION MAY CAUSE DAMAGE AND/OR FRACTURE, WHICH MAY LEAD TO A PORTION OF PRESSUREWIRE SEPARATING FROM THE TIP.
THE PRESSUREWIRE AERIS WAS BEING REMOVED FROM THE RCA WHEN THE PORTION OF THE PRESSUREWIRE DISTAL TO THE PRESSURE SENSOR BROKE OFF AND BECAME STUCK IN THE VESSEL. THE SEPARATION OCCURRED IN A PROCEDURE WHERE A NON-SJM GUIDEWIRE HAD BEEN INSERTED INTO THE VESSEL FOR PCI. THE TIP WAS REMOVED FROM THE PATIENT USING A SNARE. IT WAS REPORTED THAT THE PRESSUREWIRE WAS IN A VERY SMALL SIDE VESSEL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 696263 | PRESSURE WIRE AERIS G8 | TRANSDUCER, PRESSURE,CATHETER TIP | DXO | ST. JUDE MEDICAL, COSTA RICA LTDA | C12058 | 5425379 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |