FDA Adverse Event Malfunction Summary report: N

5MM, 33CM PEEK MONOPOLAR HANDLE 33CM

MDR report key: 6045371 · Received October 20, 2016

Report

Report Number
0002936485-2016-00988
Event Type
Malfunction
Date Received
October 20, 2016
Date of Event
September 26, 2016
Report Date
March 14, 2017
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
OCZ
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE MANUFACTURE DATE IS NOT KNOWN AT THIS TIME. HOWEVER, SHOULD IT BECOME AVAILABLE IT WILL BE PROVIDED IN FUTURE REPORTS. ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

(B)(4). ALLEGED FAILURE: DIDN'T PASS INSULATION TEST. THE FAILURE(S) IDENTIFIED IN THE INVESTIGATION IS CONSISTENT WITH THE COMPLAINT RECORD. THE PROBABLE ROOT CAUSE/S COULD BE NORMAL WEAR, USER MISUSE, AND IMPROPER STERILIZATION METHODS. THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Description of Event or Problem · 1

IT WAS REPORTED THE INSULATION IS COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THE INSULATION IS COMPROMISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
697078 5MM, 33CM PEEK MONOPOLAR HANDLE 33CM ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED OCZ STRYKER ENDOSCOPY-SAN JOSE 1441219¿H

Patients

Seq Age Sex Outcome Treatment
1