PARADIGM INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2016-33372
- Event Type
- Malfunction
- Date Received
- October 20, 2016
- Date of Event
- September 27, 2016
- Report Date
- November 7, 2016
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- K040676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(4).
THE INSULIN PUMP WAS RECEIVED WITH UNRESPONSIVE ACT AND ESCAPE BUTTONS DUE TO MOISTURE DAMAGE ON THE KEYPAD TRACES. THE INSULIN PUMP HAD MINOR SCRATCHED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP KEYPAD BUTTONS ARE NOT RESPONSIVE AND ALARMED BUTTON ERROR. THE CUSTOMER'S BLOOD GLUCOSE READING WAS UNKNOWN AT THE TIME OF INCIDENT. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACK-UP PLAN PER THEIR DOCTOR'S INSTRUCTION. THE CUSTOMER WAS ALSO ADVISED THAT THE DEVICE WOULD BE REPLACED AND TO RETURN THE PRODUCT FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 694929 | PARADIGM INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-715WWP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |