FDA Adverse Event Injury Summary report: N

INTROCAN SAFETY®

MDR report key: 6042371 · Received October 19, 2016

Report

Report Number
9610825-2016-00660
Event Type
Injury
Date Received
October 19, 2016
Date of Event
October 5, 2016
Report Date
October 5, 2016
Manufacturer
B. BRAUN MELSUNGEN AG
Product Code
DQR
PMA / PMN Number
K021094
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2011009. B. BRAUN MEDICAL INC. IS SUBMITTING A SINGLE REPORT ON BEHALF OF B. BRAUN MELSUNGEN (THE MANUFACTURER), AND B. BRAUN MEDICAL INC. (THE IMPORTER). THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). THE DEVICE INVOLVED HAS BEEN RECEIVED FOR EVALUATION AND THE INVESTIGATION IS ONGOING AT THIS TIME. A FOLLOW UP WILL BE SUBMITTED WHEN THE INVESTIGATION RESULTS BECOME AVAILABLE.

Description of Event or Problem · 0

AS REPORTED BY THE USER FACILITY: "STARTED IV, GRENN HUB DISCONNECTED FROM THE CATHETER. IV STARTED SUCCESSFULLY, CATHETER HUB DISCONNECTED FROM THE CATHETER. IV CATHETER WAS LEFT INSIDE THE PATIENTS RIGHT FOREARM. PRESSURE WAS APPLIED TO THE SITE, PATIENT WENT TO O.R. FOR A SURGEON TO ATTEMPT TO REMOVE IT." SAMPLE AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). RECEIVED 1 USED (CONTAMINATED WITH BLOOD) CAPILLARY AND CANNULA HUB OF INTROCAN SAFETY PUR 18G, 1.3X32 MM-US IN OPENED PACKAGING AND 53 PIECES INTROCAN SAFETY PUR 18G, 1.3X32 MM-US IN ORIGINAL PACKAGING. NO ABNORMALITIES FOUND ON THE CANNULA. NO ABNORMALITIES FOUND ON THE 53 PIECES OF INTROCAN SAFETY PUR 18G IN ORIGINAL PACKAGING. OBSERVED THAT THE CAPILLARY LENGTH THAT STILL INTACT WITH THE HOUSING IS APPROXIMATELY 0.7 MM. APPROXIMATELY 31.3 MM LENGTH HAS BEEN BROKEN OFF. THE REMAINING CAPILLARY OF AROUND 31.3 MM IS NOT RETURNED FOR INVESTIGATION. THE CRACK SURFACE ON CAPILLARY IS NOT ABLE TO BE REPRODUCED IN THE MANUFACTURING PROCESS AND ANY BROKEN OFF CAPILLARY CAN BE DETECTED BY THE IN-LINE VISION SYSTEM IN THE MANUFACTURING PROCESS. SIMULATION AT CAPILLARY CUTTING STATION: THE SIMULATION SHOWS THAT THE CUTTING STATION WILL NOT CAUSE SUCH DEFECT AS THE COMPLAINT SAMPLE BECAUSE THE CAPILLARY AFTER CUTTING IS A CLEAN CUT SURFACE WHICH IS NOT SIMILAR TO THE COMPLAINT SAMPLE. THE CAPILLARY OF THE SIMULATION SAMPLE AFTER CUTTING PROCESS IS APPROXIMATELY 5 MM (ORIGINALLY 37 MM) FROM THE TIP COMPARE TO THE COMPLAINT SAMPLE WHICH HAS BEEN CUT OFF APPROXIMATELY 31.3 MM FROM THE TIP WHICH WAS NOT POSSIBLE TO BE CAUSED BY THE MACHINE. IF ADDITIONAL PERTINENT INFORMATION BECOMES AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
690724 INTROCAN SAFETY® I.V. SAFETY CATHETER DQR B. BRAUN MELSUNGEN AG 16D05G8392

Patients

Seq Age Sex Outcome Treatment
1 82 YR Other