FDA Adverse Event Malfunction Summary report: N

LCS 3-PEG POLY PAT TR PINS SM+

MDR report key: 6041514 · Received October 19, 2016

Report

Report Number
1818910-2016-29808
Event Type
Malfunction
Date Received
October 19, 2016
Date of Event
September 22, 2016
Report Date
October 10, 2016
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE SUBMITTED DEVICE CONFIRMED THE PIN COMPONENT IS MISSING. THE DEVICE PIN IS A PRESS FIT AND WILL SECURE THE TWO HALVES PROPERLY. THE DIAMETER OF THE HOLE IN THE METAL BACKING PLATE WAS MEASURED AND FOUND TO MEET THE SPECIFICATION LIMITS PER THE DRAWING. MEASUREMENT OF THE PIN COULD NOT BE DONE AS THE DIS-ASSEMBLED PIN WAS NOT SUPPLIED AS PART OF THE COMPLAINT. WITH THE INFORMATION AVAILABLE, THE ROOT CAUSE OF THIS COMPLAINT CAN'T BE DEFINITIVELY DETERMINED. NO EVIDENCE WAS FOUND INDICATING DESIGN OR MANUFACTURE ERROR WERE CONTRIBUTING FACTORS AND THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. MONITOR REPORTS OF DISASSEMBLED PIN COMPONENTS THROUGH (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

RETAINING PIN HOLDING THE METAL BACKING AND TRIAL FACE HAS COME OUT.

Description of Event or Problem · 1

RETAINING PIN HOLDING THE METAL BACKING AND TRIAL FACE HAS COME OUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
693159 LCS 3-PEG POLY PAT TR PINS SM+ N/A LXH DEPUY ORTHOPAEDICS, INC. B4WGR1000

Patients

Seq Age Sex Outcome Treatment
1