FDA Adverse Event Injury Summary report: N

NA

MDR report key: 6038813 · Received October 19, 2016

Report

Report Number
0001056128-2016-00137
Event Type
Injury
Date Received
October 19, 2016
Date of Event
September 20, 2016
Report Date
September 21, 2016
Manufacturer
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
Product Code
NUJ
PMA / PMN Number
K150538
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED TO STRYKER SUSTAINABILITY SOLUTIONS FOR EVALUATION. THE DEVICE WAS THROWN AWAY DURING THE CASE. AS THE DEVICE WAS NOT RETURNED FOR EVALUATION, VISUAL AND FUNCTIONAL INSPECTION WAS UNABLE TO BE PERFORMED. A REVIEW OF THE DHR COULD NOT BE PERFORMED AS THE LOT NUMBER AND/OR SERIAL NUMBER WAS NOT REPORTED. THE REPORTED EVENT COULD BE ATTRIBUTED TO: ACTIVATION OF DEVICE BEFORE THE HANDLE HAS BEEN PROPERLY LATCHED, ANCILLARY EQUIPMENT FAILURE, THUMB TRIGGER BUTTON (ACTIVATION BUTTON) NOT ENGAGED THROUGHOUT THE ENTIRE SEAL CYCLE, ELECTRICAL CONNECTIONS DAMAGED, USE OF DEVICE ON VESSELS GREATER THAN 7 MM IN THICKNESS. THE INSTRUCTIONS FOR USE (IFU) STATE: THE LF4318 INSTRUMENT IS INTENDED FOR USE ONLY WITH THE COVIDIEN FORCETRIAD ENERGY PLATFORM. USE OF THIS INSTRUMENT WITH OTHER COVIDIEN GENERATORS OR WITH GENERATORS PRODUCED BY OTHER MANUFACTURERS MAY NOT RESULT IN THE DESIRED TISSUE EFFECT, MAY RESULT IN INJURY TO THE PATIENT OR SURGICAL TEAM, OR MAY CAUSE DAMAGE TO THE INSTRUMENT. DO NOT USE THE LIGASURE SYSTEM UNLESS PROPERLY TRAINED TO USE IT IN THE SPECIFIC PROCEDURE BEING UNDERTAKEN. USE OF THIS EQUIPMENT WITHOUT SUCH TRAINING MAY RESULT IN SERIOUS UNINTENDED PATIENT INJURY. USE CAUTION DURING SURGICAL CASES IN WHICH PATIENTS EXHIBIT CERTAIN TYPES OF VASCULAR PATHOLOGY (ATHEROSCLEROSIS, ANEURYSMAL VESSELS, ETC.). FOR BEST RESULTS, APPLY THE SEAL TO UNAFFECTED VASCULATURE. DO NOT USE THIS INSTRUMENT ON VESSELS IN EXCESS OF 7 MM IN DIAMETER. SELECT THE APPROPRIATE BAR SETTING TO ACHIEVE THE DESIRED TISSUE EFFECT. THIS SETTING MAY NEED TO BE ADJUSTED DURING THE PROCEDURE. ONE GREEN BAR ¿ USE ON ISOLATED OR SMALL TISSUE BUNDLES. TWO GREEN BARS ¿ USE WITH AVERAGE TISSUE BUNDLES. THREE GREEN BARS ¿ USE ON LARGER TISSUE BUNDLES (THIS SETTING MAY SLIGHTLY INCREASE FUSION TIMES.) NOTE: THE SYSTEM DETECTS THE CONNECTED LIGASURE INSTRUMENT AND SETS THE INTENSITY SETTING TO 2 BARS IN THE DISPLAY. IF SETTINGS ARE ENTERED IN THE LIGASURE TOUCHSCREEN PRIOR TO CONNECTING THE LIGASURE INSTRUMENT, THESE SETTINGS WILL BE RESET TO 2 BARS. PLACE THE VESSEL OR TISSUE IN THE CENTER OF THE JAWS. TO AVOID INCOMPLETE SEALING, DO NOT GRASP TISSUE BEYOND THE ELECTRODE SURFACE; DO NOT PLACE TISSUE IN THE JAW HINGE. CONTACT BETWEEN AN ACTIVE INSTRUMENT ELECTRODE AND ANY METAL OBJECT (HEMOSTATS, STAPLES, CLIPS, RETRACTORS, ETC.) MAY INCREASE CURRENT FLOW AND CAN RESULT IN UNINTENDED SURGICAL EFFECTS, SUCH AS AN EFFECT AT AN UNINTENDED SITE OR INSUFFICIENT ENERGY DISPOSITION. DO NOT ACTIVATE RF ENERGY UNTIL THE HANDLE HAS BEEN PROPERLY LATCHED. ACTIVATING THE DEVICE BEFORE THIS IS DONE MAY RESULT IN IMPROPER SEALING AND MAY INCREASE THERMAL SPREAD TO TISSUE OUTSIDE THE SURGICAL SITE. DO NOT ACTIVATE THE INSTRUMENT WHILE IN CONTACT WITH OR NEAR OTHER INSTRUMENTS, INCLUDING CANNULAS. LOCALIZED BURNS TO THE PATIENT OR PHYSICIAN MAY OCCUR. DO NOT ATTEMPT TO SEAL OVER CLIPS OR STAPLES AS THIS MAY RESULT IN AN INCOMPLETE SEAL. IF THE SEAL-CYCLE COMPLETE TONE HAS NOT SOUNDED, AN OPTIMAL SEAL MAY NOT HAVE BEEN REACHED. REACTIVATE THE RF ENERGY UNTIL A SEAL-COMPLETE TONE IS HEARD. OPEN THE JAWS BY PUSHING FORWARD THE WHITE, MOVABLE HANDLE. GRASP THE INTENDED VESSEL AND/OR TISSUE IN THE CENTER OF THE JAWS. CLOSE THE HANDLE UNTIL IT CLICKS AND LATCHES IN PLACE. ACTIVATE THE INSTRUMENT BY EITHER OF THE FOLLOWING METHODS: APPROXIMATELY PRESS AND HOLD THE PURPLE ACTIVATION BUTTON ON THE BACK OF THE INSTRUMENT. APPROXIMATELY DEPRESS AND HOLD THE CORRESPONDING ROUND, PURPLE OR ORANGE PEDAL ON THE FOOTSWITCH. A CONTINUOUS TONE SOUNDS TO INDICATE THAT THE VESSEL OR TISSUE IS BEING SEALED. WHEN THE ACTIVATION CYCLE IS COMPLETE, A DOUBLE END TONE SOUNDS AND THE ENERGY PLATFORM DISCONTINUES RF OUTPUT. RELEASE THE PURPLE ACTIVATION BUTTON ON THE INSTRUMENT, OR THE PURPLE OR ORANGE FOOT PEDAL WHEN THE SEAL CYCLE IS COMPLETE. TO SEAL ADJACENT TISSUE, OVERLAP THE EDGE OF THE EXISTING SEAL. THE SECOND SEAL SHOULD BE DISTAL TO THE FIRST SEAL TO INCREASE SEAL MARGIN. FAILURE TO OVERLAP SEALS WHEN A SECOND SEAL IS DESIRED MAY COMPROMISE SEAL INTEGRITY. KEEP THE INSTRUMENT JAWS CLEAN. BUILD-UP OF ESCHAR MAY REDUCE SEALING AND/OR CUTTING EFFECTIVENESS. WIPE JAWS SURFACES AND EDGES WITH A STERILE, WATER OR SALINE SOAKED GAUZE PAD AS NEEDED. DO NOT CLEAN INSTRUMENT JAWS WITH A SCRATCH PAD. SHOULD THE DEVICE BECOME AVAILABLE FOR RETURN, THE INVESTIGATION WILL BE REOPENED. THE REPORTED EVENT WILL CONTINUE TO BE MONITORED THROUGH POST-MARKET SURVEILLANCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN OPEN LIVER RESECTION THE REPROCESSED LIGASURE DID NOT OPERATE PROPERLY. THE PATIENT EXPERIENCED UNNECESSARY BLOOD LOSS AND REQUIRED A TRANSFUSION OF TWO UNITS OF PRBCS POST-OPERATIVELY. THE PROCEDURE WAS DELAYED APPROXIMATELY 20 MINUTES IN ORDER TO RETRIEVE ANOTHER LIGASURE DEVICE, WHICH WAS ABLE TO DIVIDE THE LIVER WITH MINIMAL BLEEDING. FOLLOWING THE TRANSFUSION, THE PATIENT WAS DOING WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692093 NA ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R NUJ STRYKER SUSTAINABILITY SOLUTIONS LAKELAND LF4318 UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1