FDA Adverse Event
Malfunction
Summary report: N
TRIAGE TOTAL 5 CONTROL L1
MDR report key: 6038710
·
Received October 18, 2016
Report
- Report Number
- 2027969-2016-00655
- Event Type
- Malfunction
- Date Received
- October 18, 2016
- Date of Event
- September 20, 2016
- Report Date
- September 29, 2016
- Product Code
- JJY
- PMA / PMN Number
- K072892
- Removal / Correction Number
- 2027969-10/12/16-004-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
MDR IS BEING FILED DUE TO FIELD ACTION REF-2027969-10/12/16-004-C. CUSTOMER'S COMPLAINT WAS REPLICATED WITH IN-HOUSE TESTING OF THE CODE CHIP OF RETAIN LOT C3233A. CODE CHIP RANGES DID NOT MATCH EV CARD RANGES. THE MANUFACTURING RECORDS FOR THE CONTROL LOT WERE REVIEWED AND THE LOT MET RELEASE SPECIFICATIONS. A CAPA, (B)(4), WAS INITIATED TO ADDRESS THIS ISSUE.
Description of Event or Problem · 1
THE CUSTOMER ATTEMPTED TO RUN THE TOTAL 5 CONTROL LEVEL 1 AT 2 STDEV WITH A CK-MB RESULT OUTSIDE OF THE EXPECTED RANGE LOW AT 3.6 NG/ML (2 STDEV: 3.7-6.6 NG/ML) AND A TNI RESULT WITHIN THE EXPECTED RANGE AT 0.33 NG/ML (2 STDEV: 0.32-0.62 NG/ML).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 689897 | TRIAGE TOTAL 5 CONTROL L1 | MULTI-ANALYTE CONTROLS | JJY | 88753 | C3233A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | CARDIAC TRIPLE MARKER, PN 97000HS, LOT# W61164B| TRIAGE METERPRO DOMESTIC KIT, PN 55070, SN (B)(4) |