LIFEVEST WCD 4000 SYSTEM
MDR report key: 6037420
·
Received October 18, 2016
Report
- Report Number
- 3008642652-2016-07553
- Event Type
- Malfunction
- Date Received
- October 18, 2016
- Date of Event
- September 14, 2016
- Report Date
- October 14, 2016
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK TE PAD MESSAGES/ADJUST BELT MESSAGES) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED A THERAPY ELECTRODE RECOGNITION TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO AN OPEN PULSE WIRE IN THE CABLE CONNECTING ECG "A" AND THE FRONT THERAPY ELECTRODE. THE ROOT CAUSE FOR THE OPEN WIRE WAS EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT.
Description of Event or Problem · 1
A US DISTRIBUTOR RETURNED ELECTRODE BELT SN (B)(4) AND REPORTED THAT A PATIENT WAS EXPERIENCING "CHECK THERAPY PAD' MESSAGES AND "ADJUST BELT" MESSAGES.