CURLIN 6000 CMS INFUSION PUMP

MDR report key: 6036676 · Received October 18, 2016

Report

Report Number
1722139-2016-00577
Event Type
Malfunction
Date Received
October 18, 2016
Date of Event
August 5, 2016
Report Date
August 9, 2016
Manufacturer
MOOG MEDICAL DEVICES GROUP
Product Code
FRN
PMA / PMN Number
K981816
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO MMDG AND EVALUATED FOR THE FREE FLOW COMPLAINT. THE PUMP DID PASS VOLUMETRIC TESTING AS WELL AS 40 PSI BACK PRESSURE TEST. THE 40 PSI TEST IS THE TEST THAT CHECKS FOR FREE FLOW. MMDG WAS UNABLE TO REPLICATE THE FREE FLOW EVENT OR CONFIRM THE COMPLAINT. THE PUMP IS OPERATING WITHIN PRODUCT SPECIFICATIONS. THE PUMP WILL BE RETURNED TO THE USER FACILITY.

Description of Event or Problem · 1

THE INITIAL REPORTER STATED THAT THE PUMP "IMMEDIATELY DELIVERED 5 ML WITHOUT ANY BUTTONS BEING PUSHED" DURING SET UP. THE INITIAL REPORTER STATED THAT THE PUMP HAD NOT BEEN PLACED ON THE PATIENT WHEN THE EVENT OCCURRED, AND THAT THERE WAS NO ADVERSE EVENT OR DELAY IN THERAPY. [(B)(4)]