FDA Adverse Event Malfunction Summary report: N

QC F/K-WIRES Ø0.6-3.2 F/532.001 532.010

MDR report key: 6036191 · Received October 18, 2016

Report

Report Number
8030965-2016-15206
Event Type
Malfunction
Date Received
October 18, 2016
Report Date
September 28, 2016
Manufacturer
SYNTHES OBERDORF
Product Code
HWE
PMA / PMN Number
K971544
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE COUPLING TOOL SIDE WAS WORN OUT. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR FROM USE AND SERVICING OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE QUICK COUPLING FOR K-WIRES DEVICE STOPPED WORKING SUDDENLY EVEN AFTER THE BATTERY WAS CHANGED. IT WAS FURTHER REPORTED THAT THE USER TESTED THE DEVICE THE NEXT DAY AND IT WORKED FINE EXCEPT IT EMITTED AN UNUSUAL SOUND. IT WAS UNKNOWN IF THERE WAS A DELAY TO THE PLANNED SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
687302 QC F/K-WIRES Ø0.6-3.2 F/532.001 532.010 INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE SYNTHES OBERDORF NA

Patients

Seq Age Sex Outcome Treatment
1