Report Number
9673241-2016-00697
Event Type
Injury
Date Received
October 18, 2016
Date of Event
September 22, 2016
Report Date
September 22, 2016
Manufacturer
BIOSENSE WEBSTER, INC. (JUAREZ)
Product Code
LPB
UDI-DI
10846835009200
PMA / PMN Number
P030031/S053
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO DEVICE WAS RECEIVED FOR ANALYSIS AT THE TIME OF SUBMISSION OF THE INITIAL 3500A. SINCE THE PRODUCT WAS NOT RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS COULD BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE WE WILL REOPEN THE COMPLAINT AND PERFORM THE INVESTIGATION AS APPROPRIATE. IF ADDITIONAL INFORMATION IS RECEIVED REGARDING THIS EVENT, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED TO THE FDA. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT NUMBER 17509318M HAS BEEN REVIEWED AND IT WAS VERIFIED THAT DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATIONS AND PROCEDURES. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). IT WAS REPORTED THAT A PATIENT UNDERWENT AN ABLATION PROCEDURE FOR ATRIAL FIBRILLATION WITH A THERMOCOOL SMARTTOUCH BI-DIRECTIONAL NAVIGATION CATHETER AND SUFFERED A CARDIAC TAMPONADE REQUIRING A PERICARDIOCENTESIS AND SURGICAL INTERVENTION. DURING THE PROCEDURE, A PERFORATION OF THE LEFT ATRIUM WAS NOTICED AS THE PATIENT BECAME HYPOTENSIVE. A TAMPONADE WAS CONFIRMED VIA INTRACARDIAC ECHOCARDIOGRAM. A PERICARDIOCENTESIS YIELDED A LARGE, UNSPECIFIED AMOUNT OF FLUID. THE PATIENT WAS TAKEN TO THE OPERATING ROOM TO STOP THE BLEEDING. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION AT THE TIME THE COMPLAINT WAS REPORTED. THERE IS NO INFORMATION ABOUT THE HOSPITALIZATION. THE RETURNED DEVICE WAS VISUALLY INSPECTED AND IT WAS FOUND IN NORMAL CONDITIONS. THE CATHETER WAS TESTED FOR ELECTRICAL PERFORMANCE, TEMPERATURE RESPONSE AND GENERATOR TEST AND IT WAS FOUND WITHIN SPECIFICATIONS. A DEFLECTION TEST WAS PERFORMED AND THE CATHETER PASSED. THE CATHETER WAS EVALUATED FOR EEPROM, AND THE FUNCTIONALITY OF THE SENSORS OF THE CATHETER WERE TESTED ON THE CARTO 3 SYSTEM. THE CATHETER WAS RECOGNIZED BY THE CARTO 3 SYSTEM, NO ERROR MESSAGES WERE DISPLAYED AND THE CATHETER WAS PROPERLY VISUALIZED. EEPROM DATA DEMONSTRATES THE CATHETER WAS PROPERLY CALIBRATED DURING MANUFACTURING. THE FORCE FEATURE WAS EVALUATED AND PASSED THE TEST. AN IRRIGATION TEST WAS PERFORMED AND THE CATHETER PASSED WITH NO OCCLUSION OBSERVED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THE CATHETER PASSED ALL SPECIFICATIONS. THE ROOT CAUSE OF THE CARDIAC TAMPONADE REMAINS UNKNOWN. THE INSTRUCTIONS FOR USE (IFU) STATES THAT CAREFUL CATHETER MANIPULATION MUST BE PERFORMED IN ORDER TO AVOID CARDIAC DAMAGE, PERFORATION OR TAMPONADE.

Additional Manufacturer Narrative · 1

THE BWI FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION ON (B)(4) 2016. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETED AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT AN ABLATION PROCEDURE FOR ATRIAL FIBRILLATION WITH A THERMOCOOL SMARTTOUCH BI-DIRECTIONAL NAVIGATION CATHETER AND SUFFERED A CARDIAC TAMPONADE REQUIRING A PERICARDIOCENTESIS AND SURGICAL INTERVENTION. DURING THE PROCEDURE, A PERFORATION OF THE LEFT ATRIUM WAS NOTICED AS THE PATIENT BECAME HYPOTENSIVE. A TAMPONADE WAS CONFIRMED VIA INTRACARDIAC ECHOCARDIOGRAM. A PERICARDIOCENTESIS YIELDED A LARGE, UNSPECIFIED AMOUNT OF FLUID. THE PATIENT WAS TAKEN TO THE OPERATING ROOM TO STOP THE BLEEDING. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION AT THE TIME THE COMPLAINT WAS REPORTED. THERE IS NO INFORMATION ABOUT THE HOSPITALIZATION. THE PHYSICIAN DID NOT PROVIDE A CAUSALITY OPINION FOR THE CAUSE OF THIS ADVERSE EVENT. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN CLARIFICATION TO THIS COMPLAINT. HOWEVER, NO FURTHER INFORMATION HAS BEEN MADE AVAILABLE. SINCE THIS ADVERSE EVENT REQUIRED MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE PERMANENT IMPAIRMENT OF A BODY FUNCTION OR PERMANENT DAMAGE TO A BODY STRUCTURE, IT IS TO BE CONSIDERED SERIOUS AND MDR REPORTABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
687579 THERMOCOOL® SMARTTOUCH® BI-DIRECTIONAL NAVIGATION CATHETER CARDIAC ABLATION PERCUTANEOUS CATHETER LPB BIOSENSE WEBSTER, INC. (JUAREZ) D-1327-05-S 17509318M 10846835009200

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening| R