FDA Adverse Event Death Summary report: N

EDWARDS SAPIEN XT TRANSCATHETER HEART VALVE

MDR report key: 6035313 · Received October 17, 2016

Report

Report Number
2015691-2016-03081
Event Type
Death
Date Received
October 17, 2016
Date of Event
September 23, 2016
Report Date
September 23, 2016
Manufacturer
EDWARDS LIFESCIENCES
Product Code
NPT
PMA / PMN Number
P130009
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PER THE INSTRUCTIONS FOR USE (IFU) CARDIOVASCULAR INJURY, SUCH AS PERFORATION DISSECTION OF VESSELS, VENTRICLE, MYOCARDIUM OR VALVULAR STRUCTURES, IS A KNOWN POTENTIAL COMPLICATION ASSOCIATED WITH THE TAVR PROCEDURE. ACCORDING TO THE THV TRAINING MANUALS, RISK FACTORS FOR AORTIC DISSECTION, HEMATOMA OR ANNULAR RUPTURE DURING THE TAVR PROCEDURE INCLUDE SIGNIFICANT THV OVER SIZING, SEVERELY OBLITERATED SINUSES OF VALSALVA, PORCELAIN AORTA AND/OR PRESENCE OF BULKY CALCIFICATION AND NARROW CALCIFIED STJ. IN ADDITION, ADVANCED AGE, FEMALE GENDER, SMALL BODY WEIGHT, AND STEROID DEPENDENCY CAN ALSO BE CONTRIBUTING FACTORS. THE SAPIEN VALVE RELIES ON NATIVE VALVE CALCIUM TO SECURELY ANCHOR TO THE ANNULUS. DESPITE THIS BENEFICIAL ASPECT OF CALCIUM, BULKY CALCIUM CAN INCREASE THE RISK OF CALCIFIC NODULE DISPLACEMENT INTO THE VASCULATURE, WHICH CAN LEAD TO VASCULAR INJURY. AT TIMES THE EXTENT AND DISTRIBUTION OF CALCIUM CAN IMPAIR EASE OF DELIVERY OF THE VALVE, CORRECT POSITIONING OF THE VALVE, DEPLOYMENT OF THE VALVE AND PROCEDURAL SUCCESS. IN THIS CASE, BASED ON THE LIMITED INFORMATION AVAILABLE THE ANNULAR RUPTURE WAS MOST LIKELY RELATED TO THE MECHANISMS DESCRIBED ABOVE. THE IFU AND TRAINING MANUALS HAVE BEEN REVIEWED AND NO INADEQUACIES HAVE BEEN IDENTIFIED WITH REGARDS TO WARNINGS, CONTRAINDICATIONS, AND THE DIRECTIONS/CONDITIONS FOR THE SUCCESSFUL USE OF THE DEVICE. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED AT THIS TIME.

Additional Manufacturer Narrative · 1

PLEASE REFERENCE RELATED MANUFACTURER REPORT NUMBER: 2015691-2016-03091 AND 2015691-2016-03110.

Description of Event or Problem · 1

AS REPORTED BY THE EDWARDS (B)(4) AFFILIATE, A 29MM SAPIEN XT VALVE WAS PREPPED AND DEPLOYED INSIDE THE PATIENT'S NATIVE AORTIC VALVE. FOLLOWING VALVE DEPLOYMENT THE PATIENT EXPERIENCED A PERI-ANNULAR HEMATOMA AS OBSERVED ON ECHO, FOLLOWED BY ASYSTOLE. THE PATIENT PASSED AWAY AND AN AUTOPSY LATER CONFIRMED THE PATIENT DIED FROM AN ANNULAR RUPTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
685528 EDWARDS SAPIEN XT TRANSCATHETER HEART VALVE AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED NPT EDWARDS LIFESCIENCES 9300TFX29

Patients

Seq Age Sex Outcome Treatment
1 76 YR Death