FDA Adverse Event Injury Summary report: N

MAR COR COMPLETE WATER TREATMENT SYSTEM

MDR report key: 6034695 · Received October 17, 2016

Report

Report Number
3019131-2016-00001
Event Type
Injury
Date Received
October 17, 2016
Date of Event
September 18, 2016
Report Date
October 17, 2016
Manufacturer
MAR COR PURIFICATION
Product Code
FIP
PMA / PMN Number
K945559
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MAR COR (MCP) TECHNICIAN ARRIVED ON A SUNDAY TO PERFORM A DIALYSIS LOOP DISINFECT SERVICE. NOTIFIED NURSES ON-SITE PRIOR TO STARTING THE DISINFECTION AND RECEIVED PERMISSION TO START THE MONTHLY DISINFECT. AFTER THE LOOP DISINFECTION AND DURING THE DIALYSIS LOOP RINSE OUT, THE MCP TECHNICIAN WAS PERFORMING A DISINFECTANT RESIDUAL CHECK OF ALL THE PATIENT ROOMS ON THE 6TH FLOOR. DURING THIS RISE OUT/RESIDUAL CHECK IT WAS DISCOVERED A DIALYSIS TECHNICIAN HAD STARTED A PATIENT ON DIALYSIS. THE DIALYSIS MACHINE WAS DISCONNECTED FROM THE WATER LOOP AND THE WATER LOOP PORT TESTED POSITIVE FOR DISINFECTANT. RINSE OUT OF THE WATER LOOP PORT WAS PERFORMED UNTIL THE RESIDUAL DISINFECTANT CHECK SHOWED NEGATIVE. THE CLINIC STAFF WAS NOTIFIED IMMEDIATELY OF THIS INCIDENT. MCP HAS SPOKEN WITH THE CLINIC AND HAS NOT BEEN NOTIFIED OF ANY PATIENT INJURY OR PATIENT REQUIRING MEDICAL ATTENTION.

Description of Event or Problem · 1

POTENTIAL OF EXPOSURE OF ONE DIALYSIS PATIENT TO MINNCARE. NO PATIENTS WERE HOSPITALIZED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
685380 MAR COR COMPLETE WATER TREATMENT SYSTEM SUBSYSTEM, WATER PURIFICATION FIP MAR COR PURIFICATION PVC LOOP

Patients

Seq Age Sex Outcome Treatment
1 Other MINNCARE