FDA Adverse Event Malfunction Summary report: N

TRIAGE TOX DRUG SCREEN PANEL

MDR report key: 6031967 · Received October 15, 2016

Report

Report Number
2027969-2016-00650
Event Type
Malfunction
Date Received
October 15, 2016
Date of Event
August 24, 2016
Report Date
August 25, 2016
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
DJR
PMA / PMN Number
K060791
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION CONCLUSION: INVESTIGATION PENDING.

Additional Manufacturer Narrative · 1

INVESTIGATION/CONCLUSION: RECEIVED 1 FALSE NEGATIVE BAR RESULT WHEN TESTING RETAIN LOT W61921 WITH POSITIVE TOX CALIBRATOR 1.5X. REVIEW OF THE DEVICE SCAN SHOWED AN ABNORMALLY THIN AND LONG PEAK AT THE BAR POSITION, THIS IS THE ROOT CAUSE OF THE FALSE NEGATIVE RESULT. NO ISSUES OBSERVED WITH TOX CALIBRATOR 0.5X TESTING. NO SAMPLE WAS RETURNED TO PRODUCT SUPPORT, FURTHER INVESTIGATION CANNOT BE PURSUED. REVIEWED THE BATCH RECORD FOR LOT W61921. LOT MET ALL FINAL RELEASE SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER QUESTIONING 3 RUNS OF TOX PANEL ON TRIAGE METER, EACH TIME RUN, DIFFERENT ANALYTES TESTED POSITIVE. CUSTOMER STATED A URINE SAMPLE FROM ONE PATIENT WAS TESTED 3 TIMES ON 2 TRIAGE METERS AND BOTH METERS GAVE THE SAME RESULTS EACH TIME, BUT THE RESULTS THEMSELVES WERE DIFFERENT FOR EACH RUN. THE 1ST RUN- 7 ANALYTES TESTED POSITIVE. THE 2ND RUN (TIME NOT PROVIDED)- JUST THC POSITIVE. THE 3RD RUN (TIME NOT PROVIDED)- JUST THC AND BAR POSITIVE. PATIENT WAS VISITING THE HOSPITAL FOR A POTENTIAL MEDICATION OVERDOSE. CUSTOMER WAS UNABLE TO RECALL PATIENT DATA DUE TO NO PATIENT IDENTIFIER GIVEN ON INITIAL REPORT. CALLER UNABLE TO GATHER INFORMATION ON WHICH SPECIFIC ANALYTES WERE POSITIVE ON THE FIRST TESTS RUN. AT THE TIME REPORTED, NO CONFIRMATORY TESTING WAS PERFORMED. ALTHOUGH REQUESTED NO CONFIRMATORY RESULTS WERE REPORTED. NO REPORTED ADVERSE PATIENT SEQUELA. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
683172 TRIAGE TOX DRUG SCREEN PANEL DOA DRUG SCREEN TEST DJR ALERE SAN DIEGO, INC. 94400 W61921B

Patients

Seq Age Sex Outcome Treatment
1 TRIAGE METERPRO DOMESTIC KIT SERIAL # UNKNOWN