FDA Adverse Event Malfunction Summary report: N

VISTA BASIC - IFVB

MDR report key: 6031369 · Received October 14, 2016

Report

Report Number
9610825-2016-00651
Event Type
Malfunction
Date Received
October 14, 2016
Report Date
September 20, 2016
Manufacturer
B. BRAUN MELSUNGEN AG
Product Code
FRN
PMA / PMN Number
K003029
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2011009. B. BRAUN MEDICAL INC. IS SUBMITTING A SINGLE REPORT ON BEHALF OF B. BRAUN MELSUNGEN (THE MANUFACTURER), AND B. BRAUN MEDICAL INC. (THE IMPORTER). THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). THE VOLUMETRIC ACCURACY WAS TESTED THREE TIMES AT 60ML/HR AND 1 HOUR (60ML) AND THE PUMP TESTED IN SPECIFICATIONS. BASED ON THE RESULTS OF THE INVESTIGATION, THE PUMP OPERATED AS INTENDED. IF ADDITIONAL PERTINENT INFORMATION BECOMES AVAILABLE, A FOLLOW UP WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2011009. B. BRAUN MEDICAL INC. IS SUBMITTING A SINGLE REPORT ON BEHALF OF B. BRAUN MELSUNGEN (THE MANUFACTURER), AND B. BRAUN MEDICAL INC. (THE IMPORTER). THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INTERNAL REPORT NUMBER (B)(4). THE DEVICE INVOLVED HAS BEEN RECEIVED FOR EVALUATION AND THE INVESTIGATION IS ONGOING AT THIS TIME. A FOLLOW UP WILL BE SUBMITTED WHEN THE INVESTIGATION RESULTS BECOME AVAILABLE.

Description of Event or Problem · 0

AS REPORTED BY THE USER FACILITY: REPORTS UNDER INFUSION. SET PUMP FOR 60ML PER HOUR INFUSED 4824 ML. REPORTS NO PATIENT INJURY. LIMITED INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
681595 VISTA BASIC - IFVB INFUSION PUMP FRN B. BRAUN MELSUNGEN AG UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1