FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 6031339
·
Received October 14, 2016
Report
- Report Number
- 3008642652-2016-07563
- Event Type
- Malfunction
- Date Received
- October 14, 2016
- Date of Event
- October 8, 2016
- Report Date
- October 14, 2016
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHARGE UNABLE TO CHARGE BATTERY PACK) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY CHARGER/MODEM WAS UNABLE TO RECHARGE A BATTERY PACK. THE CAUSE FOR THE FAILURE WAS CONTAMINATION ON THE BATTERY PCA. THE ROOT CAUSE FOR THE CONTAMINATION WAS INGRESS OF AN UNKNOWN LIQUID. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM.
Description of Event or Problem · 1
A US DISTRIBUTOR RETURNED BATTERY CHARGER/MODEM SN (B)(4) AND REPORTED THAT THE BATTERY CHARGER/MODEM WAS UNABLE TO CHARGE A BATTERY PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 682224 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |