FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 6031339 · Received October 14, 2016

Report

Report Number
3008642652-2016-07563
Event Type
Malfunction
Date Received
October 14, 2016
Date of Event
October 8, 2016
Report Date
October 14, 2016
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHARGE UNABLE TO CHARGE BATTERY PACK) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY CHARGER/MODEM WAS UNABLE TO RECHARGE A BATTERY PACK. THE CAUSE FOR THE FAILURE WAS CONTAMINATION ON THE BATTERY PCA. THE ROOT CAUSE FOR THE CONTAMINATION WAS INGRESS OF AN UNKNOWN LIQUID. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM.

Description of Event or Problem · 1

A US DISTRIBUTOR RETURNED BATTERY CHARGER/MODEM SN (B)(4) AND REPORTED THAT THE BATTERY CHARGER/MODEM WAS UNABLE TO CHARGE A BATTERY PACK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
682224 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000

Patients

Seq Age Sex Outcome Treatment
1