NASOGASTRIC SUMP TUBE DUAL LUMEN
Report
- Report Number
- 1018233-2016-01404
- Event Type
- Malfunction
- Date Received
- October 14, 2016
- Date of Event
- August 24, 2016
- Report Date
- October 28, 2016
- Manufacturer
- PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
- Product Code
- BSS
- PMA / PMN Number
- PRE-AMEND
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. THE INFORMATION PROVIDED BY BARD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BARD. THE DEVICE WAS NOT RETURNED.
THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE LOT NUMBER IS UNKNOWN; THEREFORE, THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED. THE INSTRUCTIONS FOR USE STATES THE FOLLOWING: "INDICATIONS FOR USE: BARD NASOGASTRIC SUMP TUBES ARE INTENDED TO BE USED FOR: DECOMPRESSION OF STOMACH BY SUCTION OR ASPIRATION OF GASTRIC CONTENTS. SHORT-TERM ADMINISTRATION OF TERM TUBE FEEDING, LAVAGE FLUID AND MEDICATIONS. CONTRAINDICATIONS: PATIENTS WITH KNOWN TAPE OR ADHESIVE ALLERGIES. WARNINGS: USE WITH CAUTION IN PATIENTS WITH A HISTORY OF HEAD TRAUMA, FACIAL TRAUMA, ESOPHAGEAL DISEASES AND PATIENTS WITH POTENTIAL FOR VOMITING. DO NOT FORCE NASOGASTRIC TUBE DURING INSERTIONS; DAMAGE TO THE NASAL PASSAGE AND MUCOSA AND BLEEDING MAY OCCUR. MEASURE INSERTION LENGTH CAREFULLY- EXCESSIVE INSERTION LENGTH OF TUBE INTO THE STOMACH MAY LEAD TO COILING AND/OR FORMATION OF TUBE-KNOT. LUBRICATE THE TUBE GENEROUSLY WITH WATER SOLUBLE LUBRICANT PRIOR TO INSERTION. DO NOT USE PETROLEUM-BASED PRODUCTS AS THEY MAY BE HARMFUL TO THE RESPIRATORY TRACT. REFLUX OF GASTRIC CONTENTS INTO THE BLUE VENT LUMEN INDICATES THAT THE SUCTION LUMEN IS OBSTRUCTED OR SUCTION IS TOO LOW. ROUTINELY CHECK FOR REFLUX IN THE BLUE VENT LUMEN AND CLEAR AS PER APPLICABLE DIRECTIONS. FAILURE TO CLEAR THE OBSTRUCTION OR CLEAR PREVENT FILTER MAY CAUSE GAS & FLUID BUILDUP IN STOMACH, ASPIRATION OF GASTRIC CONTENTS, ASPIRATION PNEUMONIA AND OTHER COMPLICATIONS. DO NOT INJECT FLUID THROUGH THE PREVENT FILTER AS THIS MAY RESULT IN BLOCKAGE AND LEAKAGE OF FILTER. MONITOR PATIENT FOR NASAL EROSION, SINUSITIS, ESOPHAGITIS, ESOPHAGOTRACHEAL FISTULA, GASTRIC EROSION AND PULMONARY & ORAL INFECTIONS. STATLOCK NASOGASTRIC STABILIZATION DEVICE: AVOID CONTACT WITH ALCOHOL OR ACETONE; BOTH CAN WEAKEN BONDING OF COMPONENTS AND STATLOCK STABILIZATION DEVICE PAD ADHERENCE. INSTRUCTIONS FOR NASOGASTRIC TUBE INSERTION: EXPLAIN THE PROCEDURE TO THE PATIENT. CAREFULLY MEASURE TO FIND DESIRED LENGTH OF THE TUBE USING THE NASOGASTRIC TUBE AS A MEASUREMENT AID. TO DETERMINE THE INSERTION LENGTH: MEASURE THE TUBE FROM THE TIP OF THE NOSE TO THE EARLOBE AND FROM THE EARLOBE TO THE TIP OF THE XIPHOID PROCESS. MARK THE LENGTH OF THE TUBE TO BE PASSED WITH A SMALL PIECE OF TAPE. CHECK THE PATIENT¿S NOSTRILS FOR PATENCY; SELECT THE NOSTRIL WITH BEST PATENCY. LUBRICATE THE FULL LENGTH OF TUBE TO BE INSERTED. INSERT THE TUBE THROUGH THE NOSE AIMING DOWN AND BACK. WHEN THE TUBE HITS THE PHARYNX, IF PATIENT IS ABLE, HAVE HIM OR HER FLEX HIS/HER HEAD FORWARD AND SWALLOW. ADVANCE THE TUBE AS THE PATIENT SWALLOWS. IF RESISTANCE IS MET, ROTATING THE TUBE MAY FACILITATE PLACEMENT. CONTINUE TO ADVANCE THE TUBE UNTIL THE MARKED POSITION ON THE TUBE IS REACHED. DO NOT ADVANCE BEYOND THE MARKED LENGTH AS COILING AND OR KNOTTING OF THE TUBE IN THE STOMACH MAY OCCUR. CONFIRM TUBE PLACEMENT PER HOSPITAL POLICY. THE TUBE HAS A RADIOPAQUE STRIPE FACILITATING X-RAY CONFIRMATION. IF PROPER PLACEMENT OF TUBE WITHIN THE STOMACH CANNOT BE CONFIRMED, REMOVE THE TUBE GENTLY AND START THE PROCEDURE AGAIN. SECURE WITH A SECUREMENT DEVICE OR TAPE PER HOSPITAL PROTOCOL. ENSURE 5-IN-1 ADAPTER OR LOPEZ VALVE IS SNUGLY INSERTED INTO SUCTION LUMEN TO PREVENT SUCTION LOSS. KEEP BLUE VENT LUMEN ABOVE THE LEVEL OF THE PATIENT¿S STOMACH TO PREVENT REFLUX OF STOMACH FLUIDS INTO THE BLUE LUMEN. DO NOT CLAMP AIR VENT PORT WHILE SUCTION IS BEING APPLIED." (B)(4). THE INFORMATION PROVIDED BY BARD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BARD.
THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. THE DEVICE WAS NOT RETURNED.
IT WAS REPORTED THAT THE NASOGASTRIC TUBE BROKE. THE COMPLAINANT ALLEGED THAT THE NASOGASTRIC TUBE WAS PLACED IN THE OPERATING ROOM AND WAS HOOKED TO LOW CONTINUOUS WALL SUCTION. THE PATIENT PULLED ON THE TUBE AND AS A RESULT THE TUBE BROKE WHILE INSIDE THE NASAL CAVITY. HALF OF THE TUBE WAS PULLED OUT OF THE NASAL CAVITY. THE COMPLAINANT ALLEGED THAT THE ENT WAS CALLED AND THE TUBE WAS RETRIEVED FROM THE BACK OF THE PATIENT'S THROAT AT THE BEDSIDE. ANOTHER TUBE WAS INSERTED WITHOUT INCIDENT. IT WAS REPORTED THAT NO MANIPULATION WAS DONE ON THE NASOGASTRIC TUBE PRIOR TO THE BREAK. NO PATIENT INJURY WAS REPORTED.
IT WAS REPORTED THAT THE NASOGASTRIC TUBE BROKE. THE COMPLAINANT STATED THAT THE NASOGASTRIC TUBE WAS PLACED IN THE OPERATING ROOM AND WAS HOOKED TO LOW CONTINUOUS WALL SUCTION. THE PATIENT PULLED ON THE TUBE AND AS A RESULT THE TUBE BROKE WHILE INSIDE OF THE NASAL CAVITY. HALF OF THE TUBE WAS PULLED OUT OF THE NASAL CAVITY. THE COMPLAINANT ALLEGED THAT THE ENT WAS CALLED AND THE TUBE WAS RETRIEVED FROM THE BACK OF THE PATIENT'S THROAT AT THE BEDSIDE. ANOTHER TUBE WAS INSERTED WITHOUT INCIDENT. IT WAS REPORTED THAT NO MANIPULATION WAS DONE ON THE NASOGASTRIC TUBE PRIOR TO THE BREAK. NO PATIENT INJURY WAS REPORTED.
IT WAS REPORTED THAT THE NASOGASTRIC TUBE BROKE. THE COMPLAINANT STATED THAT THE NASOGASTRIC TUBE WAS PLACED IN THE OPERATING ROOM AND WAS HOOKED TO LOW CONTINUOUS WALL SUCTION. THE PATIENT PULLED ON THE TUBE AND AS A RESULT THE TUBE BROKE WHILE INSIDE OF THE NASAL CAVITY. HALF OF THE TUBE WAS PULLED OUT OF THE NASAL CAVITY. THE COMPLAINANT ALLEGED THAT THE ENT WAS CALLED AND THE TUBE WAS RETRIEVED FROM THE BACK OF THE PATIENT'S THROAT AT THE BEDSIDE. ANOTHER TUBE WAS INSERTED WITHOUT INCIDENT. IT WAS REPORTED THAT NO MANIPULATION WAS DONE ON THE NASOGASTRIC TUBE PRIOR TO THE BREAK. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 682046 | NASOGASTRIC SUMP TUBE DUAL LUMEN | NASOGASTRIC TUBE | BSS | PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Required Intervention |